background image

33

32

Setting Markers

According to actual life status, you can set meal markers with test result.

You have to perform meal marker setting immediately after the meter displays the reading of your test result.
1. After test, the test result is displayed on the screen with time, date and flashing “            “.
2. Press the Right or Left button to choose among “            “  (before meal), “         “ (after meal), ”         ” (no marker) and   
    “             “. Then press the Main button to confirm and save this record.

Before meal: You execute this blood glucose test before a meal.

After meal: You execute this blood glucose test after a meal.
(Blank) no marker: regular test or definition by user.
No average: You will exclude this test result from average calculation. 

Please refer to "Excluding test results from the average calculation "on page 45 for more information.

NOTE

If no button pressed, the marker will be set blank automatically after 15 seconds.

Pressing for

a few seconds

Releasing 

button 

Alternative site testing: palm or forearm blood sampling 

- To perform alternative site testing, install the clear cap for your 

( For detailed information, refer to the instruction manual for the

 

. )

- To increase the blood flow, massage the intended puncture area of your

 

palm or forearm for a few 

seconds. 

-

 

Immediately after massaging the intended puncture area, press and hold the lancing device with the 

clear cap against palm or forearm.   

- Press the release button. 

- Continue holding the lancing device against your palm or forearm and gradually increase pressure 

for a few seconds until the blood sample size is sufficient.  DO NOT test on the palm or forearm if you 

are testing for hypoglycemia (Low blood glucose).

Rightest

 GD500 Lancing Device

Rightest

 GD500 Lancing Device

(

Palm

Forearm

Alternative Site Testing

All manuals and user guides at all-guides.com

Summary of Contents for GM232B

Page 1: ...40242 Taiwan R O C Made in Taiwan USA Service Center BIONIME USA CORPORATION 1450 E Spruce Street Bldg B Ontario CA 91761 USA Tel 888 481 8485 www gediabetes com Rightest blood glucose monitoring syst...

Page 2: ...All manuals and user guides at all guides com...

Page 3: ...sults to help you better manage your diabetes The Rightest Blood Glucose Monitoring System GM232B is designed for in vitro diagnostic use only for self testing by a single user outside the body Testin...

Page 4: ...or diagnose diabetes mellitus If the Rightest Blood Glucose Meter GM232B and Test Strips are exposed to a substantial change in temperature please wait 45 minutes before measurement Follow all environ...

Page 5: ...M232B and Lancing Device are for single patient use Do not share them with anyone including other family members All parts of the kit are considered biohazardous and can potentially transmit infectiou...

Page 6: ...ding Test Results and Messages Rightest Rightest Rightest Rightest Quality Control About Quality Control Testing Performing a Quality Control Test Understanding Control Test Results Test Memory Recall...

Page 7: ...Package Insert 5 Blood Glucose Meter GM232B with 2 AAA batteries installed 6 Blood Glucose Test strips GM232B 10 25 pcs 7 8 Lancing Device 9 Clear Cap 10 Disposable Sterile Lancets 10 pcs 11 Carrying...

Page 8: ...Window Test Strip Port Right Button Product name Battery Cover Meter Serial Number Meter Battery The Rightest Blood Glucose Meter GM232B The Rightest Blood Glucose Meter GM232B Rightest GM232B All man...

Page 9: ...n an error occurs Current date under time mode or testing date under memory mode Warns when the operational temperature limit is exceeded during testing Indicates the time in 12H format Displays curre...

Page 10: ...When you open a new vial of Blood Glucose Test Strips record the date on the vial Discard the vial of test strips after 3 months from opening Store the Blood Glucose Test Strip between 39 86 F or 4 3...

Page 11: ...scard a used battery Risk of explosion if battery is replaced by an incorrect type 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setting Mode Rightest all symbols will a...

Page 12: ...ain button to turn on the Rightest Blood Glucose Meter Hold down the main button for 7 seconds During this time the screen will go blank until you hear a beep After the beep the meter will turn on int...

Page 13: ...confirm Once confirmed the Hour Setting will appear Once the hour is confirmed the Minute Setting will appear Once the minute is confirmed the Volume Setting will appear 5 6 4a 4b NOTE If you do not...

Page 14: ...the Bluetooth In Time mode press and hold the Right or Left button for two seconds The blinking symbol will switch to the symbol with a beep If Volume is turned on 3 Manual turn on the Bluetooth In T...

Page 15: ...and firmly stops Removing the Rightest Blood Glucose Test Strip GM232B 1 Hold the Rightest Blood Glucose Test Strip GM232B as shown 2 Rotate the Blood Glucose Test Strip counterclockwise and pull up...

Page 16: ...e holding the base in the other hand Lift the cap until a gap appears between the cap and base 2 Pull the two parts in opposite directions to remove the adjustable depth cap 3 Insert a new disposable...

Page 17: ...The will automatically detect the Code number on the strip You do not have to check the Code number on the meter display and strip vial Rightest Blood Glucose Meter GM232B 12 A blood drop icon will ap...

Page 18: ...oo soon If this occurs please repeat the test with a new test strip Record the opening date of a new test strip vial Discard the vial of test strips 3 months after opening Always keep the metal contac...

Page 19: ...ion NOTE If no button pressed the marker will be set blank automatically after 15 seconds Pressing for a few seconds Releasing button Alternative site testing palm or forearm blood sampling To perform...

Page 20: ...ystem that is known to be well maintained and monitored by a healthcare provider use the clear cap provided with the Rightest GD500 Lancing Device when testing Make sure your blood sample covers the w...

Page 21: ...symptoms that are not consistent with your blood glucose test results and you have made sure to follow all instructions in this manual contact your healthcare professional immediately GM232B GM232B R...

Page 22: ...er control solutions please contact the GE Diabetes Customer Support Center Rightest Control Solution GC550 Level 4 Level 2 toll free at 1 866 613 7085 When should a Quality Control Test Be Performed...

Page 23: ...st Strip from the vial and close the vial cap immediately 2 Insert the Blood Glucose Test Strip horizontally not vertically with the colored view window facing up into the test strip port 3 While the...

Page 24: ...error message and Do not drip the control solution on to the sample port of the test strip directly The reagent on the strip might leak into the bottle of control solution and may cause the degenerati...

Page 25: ...the test was performed using the The control solution test result will not be used when recalling test average results Rightest Control Solution GC550 1 Press the Main button to switch the screen to...

Page 26: ...Mode you will not be able to perform the exclusion The Lo Hi results control solution test results and test results under abnormal temperature 50 F 10 C 104 F 40 C will be excluded from average calcu...

Page 27: ...and lancing device should be cleaned before being disinfected CaviWipes Disinfecting Towelettes has been tested and may be used to clean and disinfect the meter and lancing device Users may purchase t...

Page 28: ...the typical life of the meter 5 years 2 Allow to remain wet for 2 minutes 1 CAUTION Users should wash their hands thoroughly with soap and water after handling the meter lancing device or test strips...

Page 29: ...er will not function and the symbol will blink 3 If the Rightest Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature please wait 45 minutes before measurement GM232B...

Page 30: ...atteries 2 Wait for 5 minutes and reinsert the batteries as described as instructed on page 16 Meter Activation and Changing the Battery The meter should work normally after finishing the above steps...

Page 31: ...ately low results High concentrations of Ascorbic acid Vitamin C 5 mg dL Cholesterol 600mg dL and Uric acid 10 mg dL may result in inaccurate test results Not for neonatal use For over the counter use...

Page 32: ...been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interferenc...

Page 33: ...s 50 104 F 10 40 C 10 90 500 blood glucose test results with date and time Turns off automatically 2 minutes after last user action To turn off manually press themain button for 3 seconds This device...

Page 34: ...warranty card GE is a trademark of General Electric Company and is used under license by Bionime Corporation No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan This warranty does not apply...

Page 35: ...tomer service toll free at 888 481 8485 We will make every effort to provide assistance and solutions for you Description of used symbols References 1 American Diabetes Association Standard of Medical...

Page 36: ...6833 Klaus Vlbg Austria Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou JiangSu 215133 P R China Product complies with Medical Device Directiv...

Page 37: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Page 38: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Page 39: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Page 40: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Page 41: ...Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucose Insulin Me...

Page 42: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Page 43: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Page 44: ...S S Blood Glucose nsulin Medication I Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Warranty Card Distributor name...

Page 45: ...r GM232B 3 The sole obligation of Bionime under this warranty shall be to replace any defective Blood Glucose Meter GM232B No other warranties express or implied are made Bionime shall not be responsi...

Reviews: