background image

101-3GS55

1

-

080

EN

Accuracy

The accuracy of the 

Blood Glucose Monitoring System was tested by comparing 

RIGHTEST ALPHA

 

fingertip whole blood (plasma equivalent) glucose values measured by the 

Blood 

RIGHTEST ALPHA

 

Glucose Meter with plasma glucose values obtained from a YSI 2300 reference instrument.
The YSI 2300 was calibrated with NIST (SRM) 917c reference.
The results and variations between the two methods, 

 Blood Glucose Monitoring 

RIGHTEST ALPHA

System and YSI 2300 (as the reference method) are shown in the tables below.

Additional Information for Healthcare Professionals

(ii) Control solution: 

P-01

300

47.3 (2.6)
1.5 (0.08)

3.2%

P-02

300

105.6 (5.9)

2.0 (0.11)

1.9%

P-03

300

139.5 (7.7)

2.8 (0.15)

2.0%

(i) Venous whole blood sample:

Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)

Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)

(2)

Detection Principle 

The glucose oxidase and potassium ferricyanide in the strip react with the glucose in the sample to 
produce an electrical current which is proportional to the amount of glucose in the sample. The meter 
measures the current and converts it to the corresponding glucose concentration.

Performance Characteristics
Precision

The precision was evaluated by including (i) venous whole blood samples - the blood samples were 
collected over a span of time so as not to exceed one day per meter and reagent lot combination and 
(ii) 3 control solutions of different glucose concentrations (across a period of 10 days with 10 meters and 
3 batches of strips).

P-04

300

234.9 (13.0)

4.3 (0.24)

1.8%

P-05

300

374.0 (20.8)

5.5 (0.31)

1.5%

CS-L

300

46.7 (2.6)
1.2 (0.06)

2.5%

CS-N

300

98.7 (5.5)
2.0 (0.11)

2.0%

CS-H

300

289.9 (16.1)

4.1 (0.23)

1.4%

Storage and Handling

- Store the strips in the original capped vial at temperatures between 4°C to 30°C (39°F to 86°F) and 10 

to 90% relative humidity. Do not freeze.

- Replace the vial cap immediately and close tightly after taking test strip out from the vial. Do not leave      

the cap of vial opened. If the strip is exposed to the air too long, it will absorb the moisture and cause 
wrong test result.

- Every time when you open a new vial of test strips, please write the opening date on the label. Use test 

strips within 3 months after opened or until the expiration date printed on the label (whichever comes first).

Measurement Range

The measurement range of the RIGHTEST System is 10 to 600 mg/dL or 0.6 to 33.3 mmol/L.

Quality Control Section

Please refer to the Quality Control section of the User's Manual.

Troubleshooting and Customer Service

For more information on error messages and trouble shooting, please refer to the Error Messages and 
Trouble Shooting section of the RIGHTEST

 

User's Manual.

If you have any questions or in case of problems with the RIGHTEST

 

products, please contact local  

Bionime distributor or email to info@bionime.com .

NOTE

- Suggest not to use this meter close to source of strong electromagnetic radiation, to avoid 

interference with proper operation.

- Suggest to keep meter free of dust, water or any liquid.

- Severe dehydration and excessive water loss may cause inaccurately low results.
- RIGHTEST Blood Glucose Monitoring System has not been validated for use on neonates. DO NOT 

use it test for neonates.

- Do not perform the blood glucose test at temperatures below 10°C (50°F) or above 40°C (104°F), below 

10% or above 90% relative humidity. The suggested temperature range for the control solution test is 
15 - 40°C (59 - 104°F).

Difference range in values
between the YSI value and
the 

RIGHTEST ALPHA BGMS

The percent (and number) of samples was the difference between the 

RIGHTEST ALPHA BGMS

 and the YSI value within the following intervals.

Difference range in values
between the YSI value and the 

RIGHTEST ALPHA BGMS

Within ±   5%

Within ± 10%

Within ± 15%

The percent (and number) of samples was the difference between the 

RIGHTEST ALPHA BGMS

 and the YSI value within the following intervals.

*Acceptance criteria in ISO15197 : 2013 are that 95% of all differences in glucose values should be   
within 

±

 15 mg/dL (0.83 mmol/L) at glucose concentrations < 100 mg/dL (5.55 mmol/L), and within 

± 15% at glucose concentrations 

 100 mg/dL (5.55 mmol/L).

Note:

 For glucose concentrations < 100 mg/dL (5.55 mmol/L), difference values are expressed in 

mg/dL (mmol/L), and for glucose concentrations 

 100 mg/dL (5.55 mmol/L), difference values are 

compared in percentage.

Lay User Evaluation 

A total of 108 users were enrolled. Each user tested their  fingertip blood samples with 3 lots of strip 
ALPHA and ALPHA meter. Then the professional collected blood samples were centrifuged immediately 
after collection to obtain plasma. Analyze the plasma by the lab instrument (YSI 2300 analyzer). 100% of 
the ALPHA BGMS values were within ± 15% of YSI values at glucose concentrations 

 100 mg/dL (5.55 

mmol/L) and within ± 15 mg/dL (0.83 mmol/L) at glucose concentrations < 100 mg/dL (5.55 mmol/L).

Hematocrit (Hct)

Hematocrit (Hct) should be between 30 - 57%. If you do not know your hematocrit, ask your healthcare 
professional.

Interferences

26 substances were tested in low and high glucose concentrations using the blood glucose measuring 
device. Only 3 substances may interfere with glucose measurement:
Ascorbic acid 

 6 mg/dL (0.34 mmol/L)  

Glutathione reduced 

 70 mg/dL (2.28 mmol/L)

Uric Acid 

 16 mg/dL (0.95 mmol/L)

Other 23 substances within specified concentration may not interfere with glucose measurement:
Acetaminophen 

 20 mg/dL (1.32 mmol/L) Dopamine 

 2.5 mg/dL (0.13 mmol/L)

EDTA 

 0.1 mg/dL (0.003 mmol/L); Gentisic Acid 

 7.5 mg/dL (0.49 mmol/L); Heparin 

 18.75 U/mL; 

Ibuprofen 

 50 mg/dL (2.42 mmol/L); L-Dopa 

 3 mg/dL (0.15 mmol/L); Methyldopa 

 1.5 mg/dL 

(0.06 mmol/L); Pralidoxime Iodide

 ≦

 4 mg/dL (0.15 mmol/L); Salicylic Acid 

 60 mg/dL (4.34 mmol/L); 

Tetracycline 

 1.5 mg/dL (0.03 mmol/L); Tolazamide 

 15 mg/dL (0.48 mmol/L); Tolbutamide 

 

64 mg/dL (2.37 mmol/L); Bilirubin 

 50 mg/dL (0.86 mmol/L); Cholesterol 

 700 mg/dL (18.10 mmol/L); 

Creatinine 

 10 mg/dL (0.67 mmol/L); Hemoglobin 

 6000 mg/dL (0.94 mmol/L); Triglycerides 

 

3000 mg/dL (99.22 mmol/L); Maltose 

 200 mg/dL (5.55 mmol/L); Xylose

 ≦

 40 mg/dL (2.66 mmol/L); 

Galactose 

 200 mg/dL (11.10 mmol/L); Lactose 

 50 mg/dL (1.46 mmol/L); Icodextrin 

 500 mg/dL 

(30.84 mmol/L).

Reagents

Each Blood Glucose Test Strip contains the following reagents: Glucose Oxidase (GOD)               18.8%          
Potassium Ferricyanide                    37.7%

 

 

 

    Non-reactive Ingredients              43.5%

 

 

 

References

1) Diabetes Information - American Association for Clinical Chemistry(AACC)

Electronic Version

    Retrieved Jan. 26, 2021 from www.labtestsonline.org/understanding/analytes/glucose/test.html
2) 

In vitro

 Diagnostics in Diabetes: Meeting the Challenge. Clinical Chemistry 45:9, 1596-1601 (1999).             

Fingertip

72.5% (287/396)

95.5% (378/396)

100.0% (396/396)

Palm

71.2% (282/396)

95.7% (379/396)

100.0% (396/396)

Forearm

65.9% (261/396)

88.6% (351/396)

98.7% (391/396)

Fingertip

87.5% (210/240)

100.0% (240/240)

100.0% (240/240)

Palm

77.1% (185/240)

95.4% (229/240)

99.6% (239/240)

Forearm

62.9% (151/240)

96.3% (231/240)

100.0% (240/240)

Within ±   5 mg

/

dL (0.28 mmol/L)

Within ± 10 mg

/

dL (0.56 mmol/L)

Within ± 15 mg

/

dL (0.83 mmol/L)

Table 3: Represents samples for glucose results 

 100 mg/dL (5.55 mmol/L).

Table 2: Represents samples for glucose results < 100 mg/dL (5.55 mmol/L).

Table 1: Accuracy basic information

Test range in mg/dL (mmol/L)

Within ± 15 mg/dL (0.83 mmol/L) or within ± 15%

   

 

 

   

Fingertip

29 - 466 (1.61 - 25.89)

636/636 (100%)

Palm

29 - 461 (1.61 - 25.61)

635/636 (99.8%)

Forearm

32 - 463 (1.78 - 25.72)

631/636 (99.2%)

BIONIME CORPORATION

No. 100, Sec. 2, Daqing St., South Dist., 
Taichung City 40242, Taiwan
Tel: +886 4 23692388              

 

Fax: +886 4 22617586

Email: info@bionime.com        http://www.bionime.com

Emergo Europe

Prinsessegracht 20
2514 AP The Hague
The Netherlands
Email: EmergoVigilance@ul.com

Storage  tempe rature  limitation

Humidity  limitation

For 

in vitro

 diagnostic use

Manufacturer

LOT

Lot number

For single use only

EU Representive

Biological risks

CE-mark (with No. of notified body)

Consult the instruction for use

Expiry date

Importer

Rev. Date:2021-06

Reviews: