Specification
Meter Battery Life
Meter Dimensions
Meter Weight
Monitor
Display Area
Storage / Transportation Conditions
Sample
Minimum Sample Volume
Hematocrit
Test Strip Storage / Transportation Conditions
Approx. 1,000 tests
95.0 mm x 43.8 mm x 13.0 mm
43.0 ± 5 g (with battery)
LCD display
29.0 mm x 38.0 mm
10 - 40˚C (50 - 104˚F), 10 - 90% RH
Refer to RIGHTEST ELSA Strip insert
Limitations
- RIGHTEST ELSA Blood Glucose Monitoring System is not intended for serum or plasma test.
- Inaccurate test results may be obtained at high altitude more than about 3,048 meters (10,000 feet)
above sea level.
- Test results may be falsely low if the patient is severely dehydrated, in shock, or in a hyperosmolar
state (with or without ketosis).
- Patients going through oxygen therapy may yield falsely low results.- These test strips should not be
used with meters to test critically ill patients or neonates.
- The blood glucose test result may be affected by high blood concentration of interferenceingredients,
if you need more detail information about interference ingredients, please see Blood Glucose ELSA
Test Strip Insert.
- Alternative site testing (AST) should only be performed during steady-state times (when glucose isnot
changing rapidly).
- Please refer to the (Alternative site testing (AST)) chapter if you want to test on palm or forearm.
- DO NOT test on alternative site (palm, forearm) if you are testing for insulin dose calculations or
forContinuous Glucose Monitoring (CGM) devices calibration, or hypoglycemia (Low blood glucose).
- To avoid potential electromagnetic or other interference, keep meter away from
electromagneticradiation sources such as X-ray or MRI.- Keep your meter away from dust, water or
any liquid.