1
0
1
-1
G
M
1
0
0
-0
1
0
4. Time format 12/24H selection
With the time format blinking, press the main
button to adjust it. Then hold the same button
for 2 seconds to confirm it and move to hours
setting.
5. Hour setting
With the hour blinking, press the main button until the current
hour appears. Then hold the same button for 2 seconds to
confirm it and move to minute setting.
6. Minute setting
With the minute blinking, press the main button
to adjust it. Then hold the same button for 2
seconds to confirm it. Meantime, it will shift to
next digit for setting. Repeat the above action
until the minute setting is completed. Then it will
move to average-day setting.
7. Average-day setting
With the number " 1, 7, 14, 30 " blinking, press
the main button for the days of the average you
would like to read. Then hold the same button
for 2 seconds to confirm it and move to
measurement unit setting.
8. Setting the unit of measurement
With milligrams per deciliter (mg/dL) or
minimole per liter (mmol/L) blinking, press the
9. Ending setting
After confirm the measurement unit, you'll hear a sound of " beep ". All the
settings are saved and completed and will return to time screen.
1. Year setting
With the year format blinking, press the main button to adjust
it. Then hold the same button for 2 seconds to confirm it.
Meantime, it will shift to next digit for setting. Repeat the above
action until the year setting is completed. Then it will move to
month setting.
2. Month setting
With the month blinking, press the main button until the
current month appears. Then hold the same button for 2
seconds to confirm it and move to day setting.
3. Day setting
With the day blinking, press the main button to adjust it. Then
hold the same button for 2 seconds to confirm it. Meantime, it
will shift to next digit for setting. Repeat the above action until
the day setting is completed. Then it will move to time format
setting.
NOTE
-
When you do not do any settings of meter for over 2 minutes, the meter
will leave setting mode and power off automatically.
-
Any time you would like to escape from the setting mode, please press
and hold the main button for over 5 seconds. Meantime, all the current
settings will be saved.
main button to switch. Then hold the same button for 2 seconds to confirm it.
PRECAUTION
- Re-cap the test strip the vial immediately after removing a test strip.
- Do not reuse test strips.
- Do not use expired test strips.
- Record the date of opening a new strip vial for the first time. Discard the
vial of strips after 3 month from opening.
- Store the test strips in a cool and dry location. Do not expose to direct
sunlight or heat.
- For detailed information, please refer to the
GS100
Test Strip
Package Insert.
Intended Use
Blood Glucose Monitoring System
Instructions for Use
The Blood Glucose Monitoring System is intended for in vitro
diagnostic use ( for self-testing ) only ( outside the body ) and
coding
is
not
required
. The testing result is calibrated to be plasma equivalent for test with
fresh capillary whole blood samples from the fingertip, palm or forearm
depends on what kind of lancing device you are using. You may consult your
healthcare professional for instructions on how to use the system correctly. Our
customer support staff is available to assist you as well.
1.
Package of Meter Kit
2.
Preface
3
*
2
*
1
( Different packages have different bundled items. Some of packages might not
include
*
items.)
The Meter
3.
Precautions
4.
-
Before using the System to test your blood glucose, please read
all the instructions and practice the tests including quality control test.
-
Please do the quality control test regularly to make sure the test results are
accurate .
-
The GM100 Blood Glucose Meter can only be used with the
GS100 Blood Glucose Test strips. The use of any other brand
strips should not be used under any circumstances. The use of other brands
of strips may produce inaccurate results.
-
The Blood Glucose Monitoring System is intended for self-testing. It
should not be used to diagnose diabetes mellitus.
-
Blood Glucose Monitoring System has not been validated for use
on neonates. Therefore, it's not intended for use of neonates.
- This meter can be only tested with capillary whole blood. Please don't test with
arterial blood.
-
Do the test at least 30 minutes after moving into a different location with
significant change in temperature.
- Dispose of used batteries properly.
- Please note the meter kit contains small parts like test strips which could result
in a choking hazard for children.
- Prevent water from entering the meter. Never immerse the meter or hold it
under running water.
- The minimum blood sample size for testing is 1.4µl :( )
Sample Size Example
1.0 1.4 2.0 3.0
µl
µl
µl
µl
4 µl
.0
The Test Strip ( GS100 Test Strip )
5.
Sample Entry
µl
Apply a drop of blood or control solution
here. The test request is only 1.4 of blood.
Electrode Contacts
Sensing signal output terminals.
Noble Metal Electrodes
Electrochemical sensor
Indication Symbol
Insert strip with indication symbol
up and toward meter.
Battery Installation
6.
10. While the blood drop symbol flashing, you are ready to apply the blood
sample.
11. Place the lancing device against the pad of
your fingertip and press the release button.
The best puncture sites are on the middle or
ring fingers. Press the release button.
14. Pull off the depth adjustable cap. Without touching the used disposable
lancet, stick the lancet tip into the protective cover.
15. Hold the release button in one hand and pull on the plunger in the other
hand will safely eject the used disposable lancet.
16. Discard the used disposable lancet into an appropriate puncture-proof or
biohazard container.
17. Replace the depth adjustable cap after finishing the test.
12. Touch and hold the drop to the edge of
sample entry until you hear a " beep " and the
view window is totally filled with blood. If the
view window is not totally filled with blood or
the test does not start. Please discard the test
strip and repeat the test with a new test strip.
13. You will see the countdown mode on the
screen. After 8 seconds, the test result
appears.
1
2
The Blood Glucose Monitoring System was manufactured by
supported by Bionime Corporation and its authorized representative. If you
have any questions or concerns, please contact your Bionime authorized
representative or call us by phone number, 886 4 24951268 ( GMT+08:00,
08:30~17:30 ). We will provide best assistance and solutions for you.
Hand Bar
Grip here to insert
test strip into meter.
View Window
This window is yellow before applying
blood sample. It gradually turns red
when filled with blood.
We suggest you use a sample between 1.4~2.5µl. A blood sample size above
4.0µl is too large. A blood sample size below 1.4µl may give an inaccurate test
result. In this case, repeat the test with a new test strip.
GM100
is designed to use with
GS100
strip only. Please note that
misuse of other strips might cause unexpected damage or produce
inaccurate test results.
Thank you for selecting the
Blood Glucose Monitoring System.
Please read this manual thoroughly before you start testing. It provides all the
information you need to use the product. You will get a correct test result by the
following instruction of this manual.
Please only purchase the test strips in
your country, the use of the test strips of different countries is not
suggested and might get deviated test results under such circumstance.
It is quite important for you to monitor blood glucose regularly. A derived
complication can thus be effectively reduced. Through the assistance of
Blood Glucose Monitoring system, which provides considerate
design and accurate results, your diabetes management can become more
reliable and ease-to-use.
Display Window
Show your test result,
messages and relative
information.
Test Strip Port
Insert test strip here until you
hear a click.
Main Button
Press to turn on/off meter.
Switch different function
screen.
Press and hold down until
enter to the setting mode.
(There will be a power-off
period during this process.)
Battery Cover
Slide cover to
change battery.
Meter Serial
Number
Model Number
Plastic Tape
Pull it off to activate meter.
Test Strips
2007
-
06
1. Meter
2. Test Strips (0/10/25 pcs)
*
3. Control Solutions
*
4. Instructions for Use
5. Blood Glucose Test Strip
Package Insert
*
6. Lancing Device
*
7. Disposable Sterile Lancets
(10pcs)
*
8. Instruction for the lancing device
*
9. Log Book
10. Warranty Card + Emergency
11. CR2032 Battery (Inserted into the
meter)
12. Carrying Case
*
7
*
8
*
6
*
4
5
*
9. Insert the strip into the strip port on meter with the indication symbol facing
up.
11
10
13
16
14
15
17
Setting Up Your Meter-Setting the Date, Time, and Unit
……
7.
PRECAUTION
-
Danger of explosion might happen if the battery is incorrectly replaced.
-
Please follow the local regulation and discard a used battery properly.
6. Set the time and date when the batteries are replaced. See Setting the Date,
Time and Unit section. Prior test results are still stored in the memory.
Your meter comes with one CR2032, 3volt, battery installed. One
new battery will provide power to perform about 1000 tests under
normal use. Before using it, please pull off the plastic tape and
press the main button to activate the meter. The low battery
symbol " " will keep appearing on your display if your battery
runs low. Please keep spares on hand and replace the battery
soon when a battery symbol appears.
1. Turn the meter over. Press and push battery
cover to open.
2. Install the battery. Be sure to put battery in
correct direction.
3. Slide the battery cover back until it snaps into place.
4. The meter performs a self-test and all symbols on the display
will blink.
5. Press any button to exit the self-test and enter Setting Mode.
5. Choose a depth of penetration by rotating the
top portion of the depth adjustable cap until
the setting depth matches the window.
Settings are based on skin type " " for soft
or thin skin; " " for average skin; " " for
thick or calloused skin.
1. Pull off the depth adjustable cap.
2. Insert a new disposable lancet firmly into lancet carrier.
3. Twist off and set aside the protective cover of the disposable lancet.
4. Replace the depth adjustable cap.
6. Hold the hub in one hand and pull on the plunger in the other hand. The
device will be cocked. Release the plunger, it will automatically move back to
its original position near the hub.
1
2
3
4
5
6
7. Wash your hands with warm
soapy water and dry thoroughly.
8. Take one strip from the vial. Re-
cap the vial cap immediately.
7
8
9
Performing a Blood Test
8.
Hub
Lancet carrier
Plunger
Release button
Depth adjustable cap
You can enter Setting Mode by below two ways.
1. Reload battery
After removing the battery, please press the main button for several times
until no signal on screen, then follow the battery installation steps to load
battery. The meter will do self test. Press the main button to close the test
and enter Setting Mode.
2. With Battery inserted
Press the main button first to turn on the meter. Then press and hold the main
button for 5 seconds (the meter will be turned off during this period, please
keep pressing) until you hear a beep, indicating you have successfully
entered Setting Mode. The display screen will show setting data.
NOTE
-
When you keep pressing the main button for 2 seconds, the display on
screen will turn off. Please don't care it. Keep pressing the main button
till entering setting data.
-
Quick press the main button allows you to change setting while hold it
for 2 seconds will confirm the setting. However, if you press and hold
the main button over 5 seconds it will escape from setting mode and
return to time screen.
Turning on /off the Meter
1. Auto Power off
The meter will power off automatically if you don't operate it over 2 minutes.
2. Manual Power off
If you want to turn off the meter, please keep pressing the main button for 2
seconds.
3. How to turn on the Power
1) Press the main button
2) Insert one piece of test strip
PRECAUTION
Do not touch your blood drop to the sample entry on the strip until you
see the " " appear. The meter is performing an internal test and will
display " " and " " if you apply blood too soon. Then you will waste
a test strip.
12