If control solution test results are out of control solution range, your                 
System may not be working properly. Repeat the quality control test. If your 
control solution results outside the range still exist, do not use the                   
System to test your blood glucose. And contact Bionime authorized distributor 
or Bionime Customer Service.

-  our control solution has expired or after 3 months since opened.
- Your test strip has expired.
- You leave the cap off the vial of test strips or control solution for a long time.
- You didn't perform the test proce

The reason your control solution results are out of the range :

Y

dure correctly. 

Malfunction of the meter

.

Performing a Quality Control Test 

12.

1. Take one test strip from vial and Re-cap the vial 

cap immediately.

2. Insert the test strip with view window, facing up, 

into test strip port.

3. While the blood drop symbol flashing, press and hold the main button for 

over 5 seconds until the "      " symbol appears.

Use with Control Solution

4. You will see blinking "      " symbol and "      " 

symbol on the screen prompting you to apply 
control solution.

5. Shake the bottle of control solution well before 

opening the cap. Then open the cap and put it 
on the table.

6. Drip a drop of control solution on the top of the 

cap.

7. Gently touch sample entry of the strip with the 

control solution on the top of the cap.

CLOSE

  8. When you hear a beep, leave the meter on the 

table while waiting for the test result. 

      The screen will display the countdown.  
  9. Tightly replace the cap on the control solution 

bottle.

10. The control solution result appears. Compare your control test result to the 

range printed on the test strip vial label.

Battery Error

1. The "       " symbol is blinking when the battery 

power is low. Please change battery as soon 
as you can. You can still do the test.

2. The "       " and "          " symbols are blinking 

when the battery is too low. Meter can not do 
the strip test. Please change the battery 
immediately.

3. After changing the battery, perform a quality control test.

Temperature Error

In order to get accurate test result, 

 testing between 

( 50~104

 ).

perform

10~40

℃         

1. When the ambient temperature is 0~9

 ( 32~48

)

 

or 

41~50

 (106~122

) the "       " warning symbol will be 

blinking, you still can do the test but the test result is only for 
reference because the test result under these ranges of 
temperature might not be correct. Repeat the test at an area 
with temperature between operating range.

 

( 10~40

 or 

50~104

 )

2. When you move from the area with temperature outside the 

operating range of test strip to another area with temperature 
inside operating range of test strip, please wait for 30 minutes 
before you do the test.

3. When the temperature is below 0

 ( 32

 ) or over 50

         

( 122

 ), Meter can not do the test and the "         " symbol will 

blink in this condition. Please move the meter to environment 
with temperature between 10~40

 ( 50~104

 ) and repeat 

the test after 30 minutes.

Sampling Error

Before "      " appears. Please don't apply the blood to the entry of 
the strip because the meter is still doing internal check. If you do 
so, the meter will show "           " and "      ".

 

Please remove the 

strip and insert a new one to do the test again.

Meter Malfunction

If you find the meter can't be switched on, please follow the steps below for 
quick inspection.
  1. Open the battery cover, take out the battery.
  2. Wait for 5 minutes and reload the new battery. 
The meter should be work normally after finishing above steps. If not, please 
contact Bionime authorized distributor or Customer Service Center.

PRECAUTION

Check the expiration date printed on the package every time you use a 

strip. Do not use expired test strips.

Use each test strip immediately after removing from the vial.

Do not reuse test strips.

Do the test at least 30 minutes after moving into a different location with 
significant change in temperature.

Place the blood drop only on the sample entry of the strip.

- Please don't drip or inject the blood sample directly by 
  syringe to the entry port of test strip. Doing this might 
  contaminate the meter or cause damage and is strongly 
  not recommended. 

The                    Meter displays results between 10 and 600 mg/dL 
( 0.6 and 33.3 mmol/L). If your test result is below 10 mg/dL ( 0.6 
mmol/L ), "      " will appear on the screen. Please repeat your test 
again by a new strip. If you still get "       " result, you should 
immediately contact your healthcare professional.

Blood glucose test results are shown on the meter as 

 or 

 , 

depending on which unit of measurement you have chosen. Consult your 
healthcare professional before making any changes to your diabetes 
medication program.
If your blood glucose result is unusually high or low, or if you question your 
testing results, repeat the test with a new test strip. You can also run a Quality 
Control Test with                    Control Solutions to check your meter and strip. If 
the test result still remains unusually high or low, contact your healthcare 
professional immediately. 
If you are experiencing symptoms that are not consistent with your blood 
glucose test results and you have made sure to follow all instructions of this 
manual, contact your healthcare professional immediately.

mg/dL

mmol/L

If your test result is above the high end of the system's detection 
range 600 mg/dL ( 33.3 mmol/L ), "      " will appear on the screen. 
Please repeat your test again by a new strip. If you still get "      " 
result, you should immediately contact your healthcare 
professional.

Recalling Test Result & Average

10.

The                    meter is able to store 10 test results with time and date 
automatically. If your meter has stored 10 results, which is the maximum 
memory of the meter, the newest test result will replace the oldest one.
To recall your test memory, start with the meter without test strip inserted.

PRECAUTION

You have to set the time and date to activate the average function.

3. The                    meter provides you several average test results. You can 

view 1-day, 7-day 14-day or 30-day average of test results depending on 
which one you've chosen in setting mode for better blood monitoring of your 
blood glucose.

4. Press the main button until the “AVG” symbol appears on the up left corner of 

the screen. The you will see 1-day, 7-day, 14-day or 30-day average of test 
results depending on which one you've chosen in setting mode.

5. The number shown on the down left corner means how many test results are 

calculated.

6. The "       " , "      " results, the control solution results, the non - average test  

results and the test result made out of normal temperature range

   ( <10

>

40

 )

 

are not calculated in the average.

2. Under the memory screen, use main button to review all previous results with 

date and time. You will see results from the most recent ( Sequence no, " 1 " ) 
to the oldest ( Sequence no, " 10 " ) on the down left  corner of the screen.

1. Press the main button to switch the screen to memory function, you will see 

mem

 " symbol on the up right corner of the screen. it will show the latest 

testing result with time and date. The latest result is with sequence number, 

    " 1".

Insufficient blood sample

Enough blood sample

View window appearance

9.

Make sure your blood sample covers the whole area of the View Window to get 
an accurate test result. Insufficient blood sample will cause inaccurate test 
result. Repeat the test with a new test strip.

PRECAUTION

Each time you open a new bottle of 
Control solution, write the discard date 
on the label. Control solution is good 
for 3 months after opening the bottle, 
or until the expiration date printed on 
the label, whichever comes first.

Exp. Date 2008-01

When should perform a quality control test ?

Before doing a blood glucose test with your meter for the first time.
When you open and start using a new test strip vial.
When your meter is dropped or splashed with liquids.
Whenever you think your test result does not consistent with the way you feel.
Whenever you want to check if your system is working properly or not.
Whenever you want to practice testing and check correct procedure.

To make sure that your blood glucose monitoring system works properly, you 
must perform quality control testing.

1. Entering Control Solution mode

    Insert the strip into the strip port on meter. While the blood drop symbol 

flashing, press and hold the main button for over 5 seconds. Then you will 
see "      " symbol blinking on the screen indicating that you've successfully 
entered the control solution mode.

2. Control Solution Test

    A control solution test ensures that the meter and test strips are functioning 

properly together. Once you enter control solution mode, you are ready to 
perform a control solution test.

Quality Control Test

11.

PRECAUTION

- Your quality control test result by control solution test will not be 

calculated for average reading but still can be recalled. The control 
solution test result will be shown with "     " symbol on the screen.

Understanding Control Test Results

Your control solution test results should fall within the control solution range. 
That means your

                     

System is working correctly.

Example of control solution range printed 
on your test 

strip vial label

.

Control Solution Range
79-107 mg/dL
4.4-5.9 mmol/L

Maintenance

Keep your meter and test strip free of dust, water or any other liquid. Store the 
meter in the carrying case when not in use. If meter is dropped or damaged, 
perform a quality control test with control solution before doing a blood glucose 
test.

Cleaning Meter 

Clean the outside of the meter with a damp cloth and mild soap/detergent. 
Keep the test strip port from getting wet.

Don't drip the control solution to sample entry of the strip directly. The 

reagent on strip might be sucked into the bottle of control 

  solution and might cause the degeneration of control 
  solution. Doing this might contaminate the meter via the 
  test strip port as well.

Don't touch the control solution. If you have touched it, 

  please clean up with water.

Before "     " and "      " appears, please don't touch the control solution 

to the sample entry on strip because the meter is still in an internal 
check. If you do so, the meter will show "          " and "     ".

Measurement Technology

Sample

Minimum Sample Volume

Specification

Oxidase Electrochemical Sensor

Capillary 

1.4 microliter

whole blood

Error Message and Trouble Shooting

13.

Er1 - 

The inserted strip has been used or 
damaged. Please use a new test strip from 
vial.

Er2 - 

Meter has malfunctioned. Do the quality 

         control test or reinstall the battery to see 
         if the meter works properly. If error screen still appears, 

contact Bionime Customer Service.

Er3 - 

Signal transmission is disrupted, repeat the test.

95.0 mm x 43.8 mm x 13.0 mm

43.0 g with batteries

LCD display

39.0 mm x 38.0 mm

-10 ~ 60

 ( 14 ~ 140

 )

Meter Dimension

Meter Weight

Monitor

Display Area

Meter Storage Conditions

Test Strip Storage Conditions

4 ~ 30

 ( 39 ~ 86

 )

,

< 90% relative 

humidity 

Limitations

-                     Blood Glucose Test strips are designed for using with capillary 

whole blood samples. Do not use serum or plasma samples.

- Inaccurate test results may be obtained at high altitude more than about 

10,000 feet ( 3048 meters ) above sea level.

- Severe dehydration and excessive water loss may cause inaccurately low 

results.

-                     Blood Glucose Monitoring has not been validated for use on 

neonates.

- The glucose test may be interfered under abnormal concentration of 

Uric acid > 9.0 mg/dL ( > 0.54 mmol/L )  
Cholesterol > 500 mg/dL ( > 13 mmol/L ) 

Customer Service

We sincerely like to provide complete, considerate services to our customers. 
Please review all the instructions to make sure you are performing the steps 
correctly. You are always welcome to contact us by call: 886 4 24951268 
( GMT+08:00, 08:30~17:30 ), or by e-mail at 

info@bionime.com

 If you have 

any question or advice.

Warranty

Bionime Corporation warrants that your    

               

 Meter will be free from defects in 

materials and workmanship for five years from the date of purchase. 
This warranty does not apply to the performance of a    

               

 Meter that has been 

altered, misused, tampered with or abused in any way.
This warranty applies only to the original purchaser of the meter.
Please complete and return the enclosed warranty card to Bionime authorized 
representative.
Different models have different specifications. Some of the models are not included with 
the warranty card. 

Parts of Critical Component

Blood Glucose Meter, Test Strip, Control Solution and Lancing Device

Manufacturer: Bionime Corp.
Product complied with In Vitro Diagnostic Medical Device Directive 98/79/EC. ( CE0197 )
EU Rep: BIONIME GmbH; Heinrich Wild Strasse 202, CH-9435 Heerbrugg / Switzerland

For in vitro diagnostic use

Temperature limitation

Manufacturer

LOT

IVD

Lot number

Use by

Do not reuse

Method of sterilization using irradiation

Hematocrit

Power Supply

Battery Life

30 - 55%

one CR2032 battery

About 1000 tests

Disposable Sterile Lancets 

Manufacturer: Vitrex Medical ApS.
Product complied with Medical Device Directive 93/42/EEC. ( CE0459 )

Measuring Range

Test Time

Memory Capacity 

Turn off automatically after 2 minutes no 
use Press the "      " button for 2 seconds.

10 blood glucose test results with date 
and time

10-600 mg/dL ( 0.6-33.3 mmol/L )

8 seconds

Power Saving

Operating Temperature

Operating Relative Humidity

10 ~ 40

 ( 50 ~ 104

℉ 

)

10 - 90%

OPEN