Alternative Site Testing

CAUTION

- Test results may vary if blood samples are taken from different sites or under certain  

conditions where glucose levels are changing rapidly such as: following a drink, a meal, 
an insulin dose or exercise. In these cases, only the  fingertip should be used.

- DO NOT test on the palm or forearm if you are testing for hypoglycemia (Low blood 

glucose).

- Fingertip samples can show the rapid change in glucose faster than palm or forearm 

samples.

- Consult your healthcare professional before sampling from your palm or forearm. 
- Since the blood flow in the forearm and palm is slower than in the fingertip, 

sites other than 

fingertip. 

- If you use the normal lancing device for your palm or forearm, the blood sample may not 

be sufficient for meter operation.

- Periodically compare the test system to a laboratory test system that is known to be well 

maintained and monitored by a healthcare provider.

use the clear 

TM

cap provided with the 

 

Rightest

 Lancing Device GD500 when testing 

Alternative Site Testing

Releasing 

button 

Alternative site testing: palm or forearm blood sampling 

- To perform alternative site testing, install the clear cap for 

TM

your 

 

Rightest

 Lancing Device GD500. ( For detailed 

information, refer to the instruction manual for the 

 

TM

Rightest

 Lancing Device GD500. )

- To increase the blood flow, massage the intended puncture 

area of your

 

palm or forearm for a few seconds. 

-

 

Immediately after massaging the intended puncture area, 

press and hold the lancing device with the clear cap 
against palm or forearm.   

- Press the release button. 

Palm

Forearm

Pressing for

a few seconds

- Continue holding the lancing device against your palm or forearm and gradually increase pressure 

for a few seconds until the blood sample size is sufficient.  DO NOT test on the palm or forearm if 
you are testing for hypoglycemia (Low blood glucose).

(

Summary of Contents for rightest GM700S

Page 1: ...Berneck Switzerland E mail info bionime ch GM700S 0197 BIONIME CORPORATION No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E...

Page 2: ......

Page 3: ...by Bionime Corporation If you have any questions or concerns please contact your local Customer Service or call us at 886 0800 371 688 GMT 08 00 08 30 17 30 Bionime Customer Thank you for selecting th...

Page 4: ...ay give inaccurate results TM The Rightest Blood Glucose Monitoring System GM700S is intended for in vitro diagnostic use TM only The Rightest Blood Glucose test results using fresh capillary whole bl...

Page 5: ...Alternative Site Testing View Window Appearance Understanding Test Results and Messages TM Rightest GM700S TM Rightest GM700S TM Rightest GM700S TM Rightest GS700 TM Rightest GS700 8 10 14 16 18 22 2...

Page 6: ...ucose Meter with 1 CR2032 batteries installed 6 Blood Glucose Test strips GS700 10 25 pcs 7 Control Solution GC700 Normal Level is provided in your starter kit 8 Lancing Device 9 Clear Cap 10 Disposab...

Page 7: ...The Blood Glucose Meter TM Rightest GM700S Right Button Battery Cover Meter Battery The Blood Glucose Meter TM Rightest GM700S Left Button Main Button Display Window Test Strip Port 2D barcode with s...

Page 8: ...esult Test result Warns when the batteries are low or must be replaced Indicates when to apply the blood sample Appears after you insert test strip into meter Appears when an error occurs Current date...

Page 9: ...est Strip GS700 record the date on the vial TM Rightest Blood Glucose Test Strip GS700 TM If the Rightest Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature please...

Page 10: ...ease follow the local regulation to properly discard a used battery 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setting Mode TM Rightest replacing the battery all symb...

Page 11: ...onds During this time the screen will go blank until you hear a beep After the beep the meter will turn on into the Setting Mode The display screen will show setting data GM700S GM700S Setting Up Your...

Page 12: ...change any settings during Meter Set up for over 2 minutes the Blood Glucose Meter will leave setting mode and power off automatically TM Rightest GM700S Setting Up Your Meter Setting the Date Time a...

Page 13: ...no strip is inserted GM700S GM700S GM700S Turning On Off the Meter Handling the Blood Glucose Test Strip TM Rightest GS700 How to handle the ood Glucose Test Strip Inserting the Blood Glucose Test St...

Page 14: ...f the test strip port Please follow your local regulations and discard used strips properly TM Rightest GS700 1 2 3 Getting Ready for Testing Before performing a blood glucose test prepare the items b...

Page 15: ...ed skin 1 2 3 4 5 6 Performing a Blood Glucose Test 7 Hold the base in one hand and pull on the plunger with the other hand The device will be cocked Release the plunger and it will automatically move...

Page 16: ...repeat with a new Blood Glucose Test Strip GS700 15 Countdown Mode will begin on the display window After 5 seconds your test result will appear TM Rightest 13 15 Performing a Blood Glucose Test 14b 1...

Page 17: ...uld wash hands thoroughly with soap and water after handling the meter lancing device and test strips TM Rightest TM TM Rightest Rightest GM700S Setting Markers According to actual life status you can...

Page 18: ...ratory test system that is known to be well maintained and monitored by a healthcare provider use the clear TM cap provided with the Rightest Lancing Device GD500 when testing Alternative Site Testing...

Page 19: ...provider for appropriate disposal of used test strips and lancets TM Rightest TM Rightest TM TM Rightest Rightest GM700S Make sure your blood sample covers the whole area of the view window to get an...

Page 20: ...t TM Rightest Control Solution Range Example of Control Solution Range printed on your test strip vial label Control Solution Range mg dL Low 28 48 Normal 107 145 High 315 426 NOTE Normal of Control S...

Page 21: ...ate 2014 06 When should a Quality Control Test Be Performed To ensure that your Blood Glucose Meter and Blood Glucose Test Strip are working properly To confirm that you are following the correct test...

Page 22: ...trol solution This could contaminate the meter via the test strip port Keep the test strip port clean and dry Clean immediately if the test strip port is stained or is overly exposed to moisture Do no...

Page 23: ...lts TM Rightest Possible reasons your Control Solution results are out of the range Your Control solution GC700 is expired or was first opened more than 3 months ago Your Blood Glucose Test Strip GS70...

Page 24: ...lts control solution test results and test results under abnormal temperature 43 F 6 C 111 F 44 C will be excluded from average calculations The Average function requires that the correct time and dat...

Page 25: ...ort is stained with blood control solution or any liquid please use a dry tissue or alcohol swab to clean it immediately Do not use anything wet to clean it Perform a quality control test to ensure th...

Page 26: ...The meter should work normally after finishing the above steps If the meter still does not work please contact your local customer service TM Rightest GM700S Bionime Strip Error TM Rightest TM Rightes...

Page 27: ...other interference keep meter out of electromagnetic radiation sources TM Rightest Inaccurate test results may be obtained at altitudes greater than 10 000 feet 3048 meters above sea level Xylose 20 m...

Page 28: ...m x 27 5 mm 14 140 F 10 60 C 90 relative humidity 39 86 F 4 30 C Measurement Technology Sample Minimum Sample Volume Measuring Range Test Time Memory Capacity Power Saving Operating Temperature Operat...

Page 29: ...ghtest GM700S TM Different models have different specifications This warranty applies only to the Rightest Blood Glucose Monitoring System GM700S other models are not covered with this warranty card T...

Page 30: ...Taichung City 40242 Taiwan Product complies with Medical Device Directive 93 42 EEC EC Rep Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou Ji...

Reviews: