About Quality Control Testing
About Quality Control Testing
CAUTION
- Each time you open a new bottle of control solution, write the expiration date on the label.
The
Control Solution GC700 is good for 3 months after opening the bottle, or
until the expiration date printed on the label, whichever comes first.
TM
Rightest
- Wipe the bottle cap with a clean tissue before tightly closing the bottle of control solution.
- Close the bottle of control solution tightly immediately after using.
- Check the expiration date before use. (Refer to the
Control Solution GC700
Package Insert).
- Keep control solution bottles out of reach of children.
TM
Rightest
Example
Exp. Date 2014-06
When should a Quality Control Test Be Performed ?
- To ensure that your
Blood Glucose Meter
and
Blood Glucose Test
Strip
are working properly.
- To confirm that you are following the correct testing procedures.
- To prepare for your initial blood glucose test.
- To check the
Blood Glucose Test Strip
when you open a new vial of strips.
-
- To practice testing.
Required Items for Quality Control Tests
To perform a quality control test, prepare the items below :
-
Blood Glucose Meter
-
Blood Glucose Test Strips
-
Control Solution GC700
TM
TM
Rightest
Rightest
TM
Rightest
TM
To check your
Rightest
Blood Glucose Meter
after it has been dropped, damaged or
exposed to liquids.
- If you suspect that your test results are inaccurate, or if your test results are not consistent with the