- Clean and disinfect the outside of the device only. Do not remove battery cover when
cleaning and disinfecting.
Cleaning and Disinfecting frequency: at least once a week.
To clean the meter:
1. Thoroughly wipe the entire surface of the meter with disinfecting
wipes listed above to clean any possible dirt, dust, blood and other
body fluids.
To disinfect the meter:
2. Take another disinfecting wipe and wipe the meter thoroughly
(Note: All blood and body fluids should be cleaned from surface
before performing the disinfecting procedure).
3. Allow the surface to remain wet for 2 minute.
4. Allow to air dry.
Maintain the Products
Maintain the Products
Maintaining
Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying
case when not in use. If your meter is dropped or damaged, perform a quality control test with the
control solution before performing a blood glucose test.
- CAVIWIPES DISINFECTING TOWELETTES, manufacturer: Metrex. It is with Isopropanol as the
active ingredient, have been shown to be safe for use with the meter and lancing device.