Limitations
Caution
- RIGHTEST Max Blood Glucose Monitoring System is not intended for serum or plasma testing.
- Do not use at altitudes greater than 10,000 feet (3,048 meters).
- Hematocrit levels below 10% may cause falsely high reading. Hematocrit levels over 70% may
cause falsely low reading. If you do not know your hematocrit level, consult your doctor or nurse.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- The blood glucose test result may be affected by high blood concentration of interference
ingredients, if you need more detail information about interference ingredients, please see
RIGHTEST Max blood glucose test strip insert.
- Not for screening or diagnosis of diabetes mellitus.
- These test strips should not be used with meters to test critically ill patients.
- Alternative site testing (AST) should only be performed during steady-state times (when glucose is
not changing rapidly), please refer to the (Alternative site testing (AST)) chapter before you
perform AST testing.
- Alternative site testing should not be used to calibrate continuous glucose monitoring systems
(CGMS).
- To avoid potential electromagnetic or other interference, keep meter away from electromagnetic
radiation sources such as X-ray or MRI.
Important Safety Notes:
- All parts of the kit are considered biohazardous and can potentially transmit infectious
diseases, even after following the cleaning and disinfecting procedures. Please refer to the
section " Maintain the products " on page 51.
- Users should wash their hands wholly with soap and water before and after touching the
meter, lancing device, or test strips.