PRODUCT SHEET
info@biopac.com
www.biopac.com
BIOPAC Hardware |
BioNomadix Series
| Page 13 - 26
Updated: 8.18.2020
BioNomadix Transducers
Pulse BioNomadix Transducer BN-PULSE-XDCR
Emitter/Detector Wavelength:
860 nm ± 60 nm
Optical LP Filter Cutoff:
800 nm
The operational range of the emitter and detector falls within the wavelength
range of 800 nm to 920 nm. The filter is placed over the receiver; the filter of
800 nm is an optical lowpass, so wavelengths longer than 800 nm will pass
thru.
Nominal Output:
20 mV (peak-peak)
Power:
10 mA drive current
Sterilizable:
Yes (contact BIOPAC for details)
Dimensions (L x W x H):
16 mm x 17 mm x 8 mm
Transducer Weight:
4.5 grams Cable: 45 cm
Interface:
Use in CH A PPG on the BioNomadix BN-PPGED or with Smart Amplifier
PPG100D
Pulse Earclip Transducer
BN-PULSEEAR-XDR
Emitter/Detector Wavelength:
890 nm (nominal maximum)
Optical Low Pass Filter Cutoff
ambient visible light filter
The transducer operates with the BioNomadix Pulse Transmitter (BN-
PPGED) and consists of a matched infrared emitter and photo-diode, which
transmits changes in infrared reflectance resulting from varying blood flow.
Wavelength:
800-1,000 nm (70% spectral response)
Nominal Output:
20 mV (peak-peak)
Power:
10 mA drive current
Sterilizable:
Yes (contact BIOPAC for details)
Dimensions (L x W x H):
16 mm x 17 mm x 8 mm
Transducer Weight:
4.5 grams Cable length: 80 cm
Interface:
Use in CH A PPG on the BioNomadix BN-PPGED or with Smart Amplifier
PPG100D
Respiration Transducer
BN-RESP-XDCR
Response:
True DC
Circumference Range:
15 cm x 150 cm (increase with a longer strap)
Dimensions:
66 mm (long) x 40 mm (wide) x 15mm (thick)
Weight:
18 grams
Sterilizable:
YES: use standard gas sterilization techniques [i.e., Ethylene Oxide (EtO)]
Variable Resistance Output:
5 - 125 KOhm
Cable:
30 cm
Interface:
BN-RSP2 CH A RSP or CHB RSP, BN-RSPEC CHA RSP, or Smart
Amplifier RSP100D