PRODUCT SHEET
info@biopac.com
www.biopac.com
BIOPAC Hardware |
BioNomadix Series
| Page 14 - 26
Updated: 8.18.2020
Clench Force Transducer
BN-CLENCH-XDCR
Pressure Range:
0 to 1.0546 Kg-f/cm^2 (0 to 15 psi)
Error Band:
± 2% full scale
Accuracy:
±25% full scale
– best fit straight line
Output:
25 mV/0.01 Kgf/cm^2 (0.176 V/psi)
Bulb Diameter:
5.8 cm
Bulb Length:
11.1 cm
Weight:
108 grams
Cable Length:
45 cm
Interface:
Use with the BN-DYNEMG Dynamometer and EMG module
Heel-Toe Strike Transducer
BN-STRIKE-XDCR
Nominal Output Range:
-1 to +1 Volt
Nominal Contact Force:
200 g to indicate heel-toe strike
Attachment:
TAPE 1, TAPE 2, vinyl electrical or duct tape
FSR Dimensions:
18.3 mm (dia) x 0.36 mm (thick) and 30 cm pigtail lead
FSR Active Area:
12.7 mm diameter
Interface:
BN-STRIKE transmitter (STRK A, STRK B)
Skin Temperature Transducer
BN-TEMP-A-XDCR
Nominal Resistance:
2252 ohm at 25° C
Maximum operating temperature:
60° C
Accuracy and Interchangeability:
0.2° C
Response Time:
1.1 sec (attached to skin)
Compatibility:
YSI series 400 temperature probes
Sterilizable:
NO (Not designed for immersion)
Cable:
30 cm
Dimensions:
9.8 mm (diameter) x 3.3 mm (high)
Interface:
BN-SKT2 only: CH A SKT and/or CH B SKT, or Smart Amplifier SKT100D
Skin Temperature Transducer
BN-TEMP-B-XDCR (Fast Response)
Nominal resistance:
2252 ohm @ 25° C
Maximum operating temperature:
60° C (when used with BN-SKT2)
Accuracy and Interchangeability:
0.2° C
Response Time:
0.6 sec (in air)
Compatibility:
YSI series 400 temperature probes
Sterilizable:
NO (Not designed for immersion)
Cable:
30 cm
Dimensions:
1.7 mm (diameter) x 5 mm (long)
Interface:
BN-SKT2 only: CH A SKT and/or CH B SKT, or Smart Amplifier SKT100D