BN-ACCL ACCELEROMETER CALIBRATION
Orient Transmitter unit in the X, Y, and Z directions with respect to Earth’s gravity. This action will introduce 1
G in positive axis direction and -1 G in the negative axis direction. The accelerometer Transmitter/Receiver set
has user-selectable ranges: ±2, ±4, ±8 or ±16 G. The maximum value of each range selection provides a +10 V
output and the minimum value of each range selection provides a -10 V output. When using the ±2 G range, a + 1
G input will provide a +5 V output and a -1 G input will provide a -5 V output, when properly scaled.
The BN-ACCL has a factory default of ±16 G. Use the DIP switches on the side of the BN-ACCL Receiver
Module to set the Transmitter to the desired range. If the range needs to be set to something other than ±16 G,
perform the following steps prior to calibration using Earth’s gravity, as described in the previous paragraph.
1.
In Acq
Knowledge
, select “
Set Up Data Acquisition
>
Channels
.”
2.
Select “
View by Modules
...” and add new module “
ACCL3-R
,”
assuming not yet added.
3.
Select the desired X, Y and Z channels and click
OK
(see right).
4.
Then select “
View by Channels
...”. Click the “
Setup
” button.
5.
Click “
Yes
” to the channel scaling modification prompt.
6.
Set desired Scaling as shown in the following table and click
OK
:
±16 G Range
±8 G Range
±4 G Range
±2 G Range