PRODUCT SHEET
info@biopac.com
www.biopac.com
BIOPAC Hardware |
BioNomadix Series
| Page 24 - 26
Updated: 8.18.2020
RSP2-R BioNomadix Receiver
*
indicates
Factory
Preset
PPGED-R BioNomadix Receiver
Filter Option
CH A
CH B
Filter Option
PPG CH A
EDA CH B
Low Pass
SW3
SW5
Low Pass
SW3
SW5
10 Hz LP
DOWN
DOWN
3 Hz LP
UP
*
UP
*
1 Hz LP
UP
*
UP
*
10 Hz LP
DOWN
DOWN
High Pass
SW4
SW6
High Pass
SW4
SW6
0.5 Hz HP
UP
UP
0.5 Hz HP
UP
*
UP
DC
DOWN
*
DOWN
*
DC
DOWN
DOWN
*
RSPEC-R BioNomadix Receiver
Filter Option
RESP CH A
ECG CH B
Low Pass
SW6
SW4
1 Hz LP
UP
*
35 Hz LP
UP
*
10 Hz LP
DOWN
150 Hz LP
DOWN
High Pass
SW7
SW3
0.5 Hz HP
UP
1 Hz HP
UP
*
DC
DOWN
*
0.05 Hz HP
DOWN
NICO-R BioNomadix Receiver
Filter Option
Switch Number
Low Pass
SW1 (Z CH)
SW2 (dZ CH)
5 Hz LP
UP
UP
Low Pass
SW3 (Z CH)
SW4 (dZ CH)
3 Hz LP
UP
UP
Low Pass
SW5 (ZCH)
SW6 (dZ CH)
1 Hz LP
UP
UP
DC to 50
*
Hz
DOWN for all switches
*
*
Bandlimited to 10 Hz in BN-NICO units shipped before 11/2016
DYNEMG BioNomadix Receiver
Filter Option
DYN CH A
EMG CH B
Low Pass
SW6
SW4
3 Hz LP
UP
250 Hz LP
UP
100 Hz LP DOWN
*
500 Hz LP
DOWN
*
SW7
SW5
10 Hz LP
UP
*
Envelope detection mode EMG CH B
UP
100 Hz LP
DOWN
EMG signal output
DOWN
*
High Pass
HP N/A for DYN
SW3
--
--
10 Hz HP
UP
*
--
--
1.0 Hz HP
DOWN