PRODUCT SHEET
info@biopac.com
www.biopac.com
BIOPAC Hardware |
BioNomadix Series
| Page 8 - 26
Updated: 8.18.2020
Table 2: BioNomadix Dual Transducer
– See Table 1 for Biopotentials, Table 3 for Combo Pairs, and Table 4 for Accelerometers
BioNomadix
BN-SKT2
BN-RSP2
BN-GONIO
BN-STRIKE
Signal type:
Dual Channel SKT
temp
Dual Channel RSP
resp
Dual Channel Goniometry
Dual Channel Strike Data
BandlimitsMax:
Factory preset:
Filter Options:
DC to 10 Hz
DC to 1 Hz
DC, 0.5 Hz HP, 1 or 10 Hz LP
DC to 10 Hz
DC to 1 Hz
DC, 0.5 Hz HP, 1 or 10 Hz LP
DC to 100 Hz
DC to 10 Hz
DC, 3 Hz, 10 Hz, or 100 Hz
LP
DC to 100 Hz
DC to 10 Hz
DC, 3 Hz, 10 Hz, or 100 Hz LP
Notch filter:
50/60 Hz user-controlled switch; typically not required
—factory preset OFF
.
See Appendix for
additional hardware-specific output options.
50/60 Hz user-controlled
switch
– factory preset OFF
Resolution:
0.01° C (rms)
FSR/4096; (4.88 mV)
0.01
rotation (rms)
N/A
Signal range:
13 to 51° C
± 10 V (at output)
± 180
± 10 V (at output)
Output Voltage range:
± 10 V (receiver output)
Operating time:
72-90 hours
Included strap:
137 cm - BN-STRAP-137
137 cm - BN-STRAP-137
76 cm - BN-STRAP-76 &
BN-STRAP-33
33 cm - BN-STRAP-33
Input:
BN-TEMP-A/B-XDCR
BN-RESP-XDCR
BN-GON-110-XDCR
BN-GON-150-XDCR
BN-GON-F-XDCR
BN-TOR-100-XDCR
BN-TOR-150-XDCR
BN-STRIKE-XDCR
BN-ADAPT-GONIO
This adapter connects goniometers and torsiometers to a wireless BN-GONIO Transmitter, use one adapter per channel.
BioNomadix goniometers and torsiometers include the required BN-ADAPT-GONIO adapter(s):
•
Two adapters included with BN-GON-110-XDCR and BN-GON-150-XDCR goniometers
•
One adapter included with BN-GON-F-XDCR finger goniometer and BN-TOR-110-XDCR and BN-TOR-150-XDCR torsiometer
Adapters can also be used with existing BIOPAC or 3rd-party goniometers and torsiometers to make them compatible with the BioNomadix wireless
transmitter. These adapters are required if not using BioNomadix Transducers.