USER MANUAL
TD064-O&M-001
REVISION 4
Page 16 of 38
Press “HPV CYCLE” if enabled;
otherwise the cycle will have to be run
from the Generator or the other door
screen.
Note: The following screens, prior to starting a cycle, will differ
dependent upon which HPV generator is connected to the WIC. When
connected to an IG-1 or L-3 the data from the RFID tag on the Hydrogen
Peroxide bottle is displayed on screen and the control system prevents
the cycle from running with invalid or insufficient Hydrogen Peroxide.
The Bioquell L-2 does not have this functionality and as such the
Operator must ensure enough Peroxide is available in the bottle.
Bioquell L-2:
Bioquell IG-1/L-3:
The cycle can be selected by pressing
on the black space and selecting the
cycle number.
To record the equipment to be
decontaminated; enter in the black
space using the keyboard which will
appear. This reference will display on
all screens and will be printed.
When setup is complete press the tick
icon. To enable the decontamination
process to commence press the yellow
confirmation button that appears when
connection setup is complete.
Press the >> button to view the
chamber in OFF mode prior to
confirming setup complete.