If concentration is OK then the display
will prompt the Operator to end the
cycle.
Press the tick icon to confirm.
It will be necessary at this stage to
enter the Operator’s password.
Screen displayed is dependent upon whether the doors are interlocked:
No door interlocks
With door interlocks
At this stage the Operator can proceed to unlock the door and remove the
contents.
4.3
Unloading the Chamber
Unlock the clean side door.
Remove the items in the Chamber, and place any items which require passing
back into the Chamber.
If the door interlock is enabled then at the end of unloading the chamber the
“Operations Complete?” button must be pressed to allow the other door to be
opened and the chamber re-loaded for the next cycle.
A decontamination cycle can be run from either side of the Chamber by amending
the door priority (if option enabled during setup).