USER MANUAL
TD064-O&M-001
REVISION 4
Page 15 of 38
During operation both Door A and Door
B status will display locked and closed
and the “UNLOCK DOOR” button is not
visible.
Count down time.
When the AIR CLEAN UP is complete
press the now visible “UNLOCK DOOR”.
Either door can be opened.
4.2.2
Running a Bio-Decontamination Cycle
Ensure the HPV Generator is connected and operational in accordance with the
instructions within the specific Generator User Manual.
When about to start a cycle check that all the connections have been made
correctly.
Check the hose carrying vapour from the Generator into the chamber is
connected, and the Camlock arm rings are secured and locked at both
ends.
Check the hose returning the vapour from the Chamber to the Generator
is connected and the Camlock arm rings are secured and locked at both
ends.
Check the communications cable plug is connected into the back of the
Generator and into the Chamber.