USER MANUAL
TD064-O&M-001
REVISION 4
Page 22 of 38
4.4
Running a Generator Self-Bio-decontamination Cycle
The HPV Generator can be transferred through the chamber and bio-
decontaminate itself by the following method:
Turn OFF and disconnect power to the Generator.
Disconnect the hoses and control cable.
Place the Generator into the chamber, suitably configure it, and connect to
power supply at the back of the Generator.
Connect the control cable to the back of the Generator and to the
connection within the Chamber adjacent to the power sockets.
Start cycle as previously described.