4
Biotage
®
Extrahera
™
Installation and Safety | © Biotage 2020
Installation
24. Connect a pressurized air or nitrogen supply to the
AIR
port
located on the right hand side at the rear of the system using
the pressure regulator as shown in Figure 7.
The gas connected to the
AIR
port is used to seal the plate/
columns. The pressure should be adjusted according to how
many positions in the extraction plate or column rack that
are populated. Use 6 ± 0.2 bar (0.6 ± 0.02 MPa; 87 ± 3 psi)
for a fully populated plate/rack and lower to approximately
4 bar (0.4 MPa; 58 psi) when 50% of the plate/rack is
populated and 3 bar (0.3 MPa; 44 psi) when 25% of the
plate/rack is populated.
Figure 7
Connect to
the
AIR
port
Connect to
supply
.
The setup of the pressure regulator for the AIR port.
25. Connect a pressurized air or nitrogen supply to the
N
2
port
located on the right hand side at the rear of the system using
the filter unit with a pressure regulator; see Figure 8. Set the
pressure to 6 ± 0.2 bar (0.6 ± 0.02 MPa; 87 ± 3 psi).
Figure 8
Connect to
the
N
2
port
Connect
to supply
.
The setup of the filter unit for the N
2
port.
26. If you have the Extrahera GLP software and you want to
connect the system to your network, connect a network
cable between the
LAN
port located on the right hand side
of the system and your network.
27. If you have disconnected the touch screen, reconnect it by
reversing the instructions in step 14.
Note:
Do not mix up the Ethernet cables. Connect according
to the labeling on the cables and the ports (
LAN 1
and
LAN 2
).
28. Connect the system to a properly grounded (earthed) power
outlet with the correct mains voltage and frequency.
29. Ensure that the power cord and any cables, hoses, and
tubing connected to the system cannot come in contact with
chemicals. Corrosives and solvents can degrade the cord/
cable insulation and dissolve the hoses and tubing. There is
a risk of electric shock, fire, and/or equipment damage.
30. Turn on the system. The mains switch is located on the right
hand side at the rear of the system.
31. Purge the system:
a. Press
Maintenance
in the software's main menu.
b. Press
Manual Control...
and then
Close
in the
Door
field.
c. Press
Move Out
in the
Pressure Unit
field.
d. Enter 0.5 bar in the
Pressure (bar)
text box; see Figure 9.
e. Press
Apply
, wait for 15 seconds and then press
Stop
.
f. Press
Move In
in the
Pressure Unit
field.
g. Press
Open Door
.
Figure 9.
The Manual Control view.
32. Reconnect the gas tubes to the pressure head. Ensure that
they are properly fastened by pulling on them and that they
are on the left side of the bracket as shown in Figure 10.
Figure 10.
Ensure that the gas tubes are on the left side of the highlighted
bracket.
33. Put the pressure head back in place using the two screws.
Do not overtighten the screws.