background image

5

Installation

34. Place the pipette tip waste bin for used pipette tips below 

the 

WASTE

 position on the working area.

35. Place the solvent rack with five new, empty solvent reservoirs 

in position 

on the working area. Ensure that the solvent 

feeder is pulled out into the correct position; see Figure 11.

Figure 11.

 The solvent feeder in position.

Connections

Figure 12

1

2

3

4

5

.

 Solvent connections located on the back wall inside the system.

Port

Label

Connect

1

S1

The S1 solvent inlet line (right port) and the 

S1 solvent feeder tube (left port).

2

S2

The S2 solvent inlet line (right port) and the 

S2 solvent feeder tube (left port).

3

S3

The S3 solvent inlet line (right port) and the 

S3 solvent feeder tube (left port).

4

S4

The S4 solvent inlet line (right port) and the 

S4 solvent feeder tube (left port).

5

S5

The S5 solvent inlet line (right port) and the 

S5 solvent feeder tube (left port).

Figure 13

6

7

8

.

 Waste connections located on the back wall inside the system.

Port

Label

Connect

6

Waste outlet tube between the waste valve 

and the waste reservoir.

7 + 8

Waste tube between the waste valve and 

the extraction waste collector.

Figure 14

10

9

12

11

.

 Connections located on the right hand side at the rear of the 

system.

Port

Label

Connect

9

T4A-250VAC Electrical supply: 100 to 240 V~, 50/60 Hz, 

300 VA

Fuses: T4A (2 required)

10

LAN

Network. Optional.

11

N

2

The gas applied to the samples through 

the pressure head. See the gas supply 

requirements in the technical specification 

on page 6.

12

AIR

The gas used for the movement of the 

pressure head. See the gas supply 

requirements in the technical specification 

on page 6

Upgrade the Biotage

®

 Extrahera

 Software

Please refer to the instructions delivered with the new software. 

Summary of Contents for Extrahera Classic

Page 1: ...Biotage Extrahera Installation and Safety...

Page 2: ...l Specifications 6 System Requirements 6 Liquid Handling 6 Interfaces 7 Clog Detection 8 Safety 8 Intended Use 8 Education Training and Competence 8 Warranty and Liability 8 Service 8 Safety Features...

Page 3: ...unded outlet Vacuum supply Partial pressure range 500 mbar 50 kPa 7 psi Flow rate 8 3 L min 0 29 cubic feet min Connection The vacuum tube supplied with the system has a 6 mm 0 24 outer diameter and 4...

Page 4: ...in position 5 on the working area c Press Maintenance in the main menu d Press Flush Solvent Inlets e Enter the flush volume for S1 25 mL is required to empty the solvent inlet line and pump of liquid...

Page 5: ...e height of the feet There is a built in spirit level in the bottom of the system in the front left corner See Figure 3 Figure 3 The built in spirit level in the bottom of the system in the front left...

Page 6: ...system and your network 27 If you have disconnected the touch screen reconnect it by reversing the instructions in step 14 Note Do not mix up the Ethernet cables Connect according to the labeling on t...

Page 7: ...and the S4 solvent feeder tube left port 5 S5 The S5 solvent inlet line right port and the S5 solvent feeder tube left port Figure 13 6 7 8 Waste connections located on the back wall inside the system...

Page 8: ...s populated and 3 bar 0 3 MPa 44 psi when 25 of the plate rack is populated N2 port 1 L min 0 04 cubic feet min during processing and 70 L min 2 5 cubic feet min during plate dry 6 0 2 bar 0 6 0 02 MP...

Page 9: ...d Heptane Good Hexane Poor Isooctane Good Isopropanol Excellent Methanol Excellent Pentane Poor Tetrahydrofuran THF Poor Toluene Excellent Water Excellent Methyl tert butyl ether MTBE Poor IPA 30 Exce...

Page 10: ...h hazardous biological radioactive or toxic samples and or solvents prior to any service being performed When returning equipment to Biotage this should be done in accordance with the material return...

Page 11: ...e suitable lifting equipment when moving the package The system weighs 75 kg 165 lbs The four lifting handles supplied with the system must be used when moving the system Four persons are required whe...

Page 12: ...ulations All samples and waste should be treated as potentially biohazardous Always follow local and national safety regulations related to storage handling and disposal of chemicals biological sample...

Page 13: ...ic Equipment WEEE Products falling under the scope of the WEEE Directive are identified with a crossed over wheelie bin symbol on the product label as indicated to the left To forward a product for re...

Page 14: ...the back of this document or visit our website www biotage com Manufacturer Biotage GB Limited United Kingdom for Biotage Sweden AB Contact Us Biotage Sweden AB Box 8 SE 751 03 Uppsala SWEDEN Visiting...

Page 15: ......

Page 16: ...Part Number 414157 J 2020 Biotage All rights reserved No material may be reproduced or published without the written permission of Biotage Information in this document is subject to change without no...

Reviews: