background image

15 / 18

Table 3:

Guidance and manufacturer’s declaration - electromagnetic Immunity

Radiated RF

IEC61000-4-3

(Test specifications

for ENCLOSURE

PORT IMMUNITY

to RF wireless

communications

equipment)

Test

Frequency

(MHz)

Band

(MHz)

Service

Modulation

Modulation

(W)

Distance

(m)

IMMUNITY

TEST LEVEL

(V/m)

385

380

–390

TETRA 400

Pulse

modulation

18 Hz

1,8

0.3

27

450

430

–470

GMRS 460,

FRS 460

FM ± 5 kHz

deviation

1 kHz sine

2

0.3

28

710

704 –

787

LTE Band

13,

17

Pulse

modulation

217 Hz

0,2

0.3

9

745
780
810

800–96

0

GSM

800/900,

TETRA

800,

iDEN 820,

CDMA 850,

LTE Band 5

Pulse

modulation

18 Hz

2

0.3

28

870

930

1720

1700 –

1990

GSM 1800;

CDMA

1900;

GSM 1900;

DECT;

LTE Band 1,

3,

4, 25;

UMTS

Pulse

modulation

217 Hz

2

0.3

28

1845

1970

2450

2400 –

2570

Bluetooth,

WLAN,

802.11

b/g/n,

RFID 2450,

LTE Band 7

Pulse

modulation

217 Hz

2

0.3

28

5240

5 100 –

5 800

WLAN

802.11

a/n

Pulse

modulation

217 Hz

0,2

0.3

9

5500

5785

Summary of Contents for LTH-1

Page 1: ...1 18 Non contact Infrared Thermometer LTH 1 User Manual Guangzhou Labsim Biotech Co Ltd ...

Page 2: ...n and storage environment 5 3 Installation and usage 5 3 1Install the battery 5 3 2How to replace the battery 6 3 3Measurement unit setting 6 4 Operation method 6 4 1Measuring body temperature 6 4 2Memory function 7 4 3 Shutdown 7 5 Product symbol description 8 6 For technical data please contact 9 7 Maintenance 9 8 Common Faults and Solution 10 9 Warning Precautions and Limitations 11 10 EMC Decl...

Page 3: ...ed for measuring heat radiation from the human forehead it can show the body temperature of human The product can be reused 1 3 Composition and Structure 1 1 1 Non contact infrared thermometer mainly consists of infrared sensor distance sensor control circuit board display screen operation buttons plastic shell probe protection cover may not include The schematic diagram of its structure is shown ...

Page 4: ... and adults 1 6 Contraindications No 2 Technical parameters and conditions of use 2 1 Technical parameters Power supply DC 3V 10 5 Measurement range 32 0 43 0 Display resolution 0 1 Measurement accuracy Within 35 0 42 0 0 2 0 3 within the range of 32 34 9 and 42 1 43 0 Shock protection internal power supply equipment Application part BF type applied part Operating mode continuous operation Tempera...

Page 5: ...nment working environment Temperature 15 to 40 Humidity 15 to 90 RH Atmospheric pressure 70kPa 106 kPa Transportation and storage environment Temperature 25 to 70 Humidity 90 RH Atmospheric pressure 70kPa 106 kPa 3 Installation and usage 3 1Install the battery This machine comes with 2 pieces of AAA batteries Push the battery cover down and insert the battery into the battery compartment At this t...

Page 6: ...tive direction of the battery and the power Note The battery with the machine is a test battery please use new batteries when the testing institution tests 3 3Measurement unit setting In the shutdown state long press the memory key for 2 3 seconds to enter the temperature unit conversion state then short press the memory key to select and temperature units then press the power on measurement key t...

Page 7: ...d to measure as shown in Figure 4 Figure 4 2 When the measurement distance is less than about 3 cm it will automatically measure with a di sound and display the measurement result 3 When the body temperature exceeds 37 5 the screen will be displayed in orange and at the same time the machine will have three rapid alarm sounds di di di 4 When the measured value exceeds the measuring range 32 43 the...

Page 8: ...e with care Time limit for environmental concerning DISPOSAL Do not dispose this product as unsorted municipal waste Collection of such waste separately for specia treatment is necessary Keep dry Manufacture Authorized representative in the European Community Serial Number Follow Instruction for use CE marked product Date of Manufacture Medical device ...

Page 9: ...pped in a little medical alcohol 75 to gently wipe the surface of the product and the measured head Use it after the alcohol has evaporated Precautions for storage This product should be stored in a dust free and dry place Please avoid direct sunlight do not store it in a place with high temperature humidity dust and corrosive gas This product is a high precision product please do not drop the dev...

Page 10: ...en display The measuring temperature is lower than 43 which exceeds the measuring range of the machine Please re measure according to the instruction screen display The ambient temperature is too low or too high Please measure at 16 35 ambient temperature screen display Low battery Please replace the battery No display or abnormal display The battery is reversed or the machine is abnormal Reinstal...

Page 11: ...ary impact on its accuracy 7 Before use please read this instruction manual carefully to confirm that the battery is installed 8 Do not disassemble repair or modify this product otherwise it may cause errors in the measurement results or cause machine failure 9 This product is not waterproof Use it with care to prevent liquid from entering the product When this product gets wet due to contact with...

Page 12: ...easuring do not face the sun air outlet of the air conditioner or heater which will change the temperature Please try to measure in a stable environment 17 Under the conditions of storage regulations the life of the product is 5 years except for vulnerable and consumable parts ...

Page 13: ...turer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the equipment including cables specified by the manufacturer O...

Page 14: ...licable Not applicable Surge IEC 61000 4 5 Not applicable Not applicable Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Not applicable Not applicable Power frequency magnetic field IEC 61000 4 8 30 A m 50Hz 60Hz 30 A m 50Hz 60Hz Conducted RF IEC61000 4 6 Not applicable Not applicable Radiated RF IEC61000 4 3 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz 1...

Page 15: ... 0 3 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0 3 9 745 780 810 800 96 0 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz 2 0 3 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 ...

Page 16: ...facturer 2 For the failure caused by the following users personal reasons free warranty service will not be provided 1 Failure caused by unauthorized disassembly and modification 2 Failure caused by accidental drop of the product 3 Failure caused by not operating according to the instructions 4 Failure caused by lack of reasonable maintenance 5 Damage caused by external force 3 Repair services out...

Page 17: ...17 18 13 warranty card warranty card Model Serial No Purchase date username contact number Seal of sales point Maintenance records matter Repairer Remarks Please show this card for repair ...

Page 18: ...ong Province PEOPLE S REPUBLIC OF CHINA Tel 86 20 28200186 Fax 86 20 28200183 After Sales Service Tel 86 400 617 6839 zip code 510700 E mail gzlb thergane com Website www labsim cn CMC Medical Devices Drugs S L Add C Horacio Lengo No 18 CP 29006 Málaga Spain Telephone 34 951214054 Email Info cmcmedicaldevices com version number A 1 Date of edit April 15th 2022 ...

Reviews: