P
AGE
20
OF
25
Connect the water supply to the Venturi Sampler. The connection is ¼” Swagelok.
The water pressure is typically set to a maximum of 3 bar. The pressure must always be
lower than the purge air pressure. If it is higher than the purge air pressure, then water can
feed into the air line.
_____
In cases where there is sticky or fatty material in the sample, it may be necessary to use hot
water to backwash the venturi sampler.
_____
Connect the sample tube (or tubes) from the Venturi Sampler to the BioTector in accordance
with the supplied drawing package.
_____
If a particulate separation chamber
“External Chamber” option is supplied, install the chamber
at the sampling point.
_____
Electrically connect the Venturi Sampler to the BioTector with the supplied cable, in
accordance with the supplied drawing package.
_____
Check the electrical connections and confirm there are no loose connections within the Venturi
Sampler.
_____
Check the Swagelok / PFA tube connections and confirm there are no loose connections within
the Venturi Sampler.
_____
Water connection
Water pressure
set to 3 bar
maximum using
customer
supplied water
regulator
BioTector
connection,
¼” OD tube