P
AGE
8
OF
25
4.
Typical Operational Data
Build parameters.
First chamber volume:
70ml.
Second chamber volume:
270ml.
Sample tubing:
½” OD x 3/8” ID.
Sample distance, 40 meters, typical filling and purging times
Lift,
meters
Time for venturi
sampler to fill tubes
and chamber
Venturi filling time,
including pre-purge
times for 4 streams.
Add 8s for 6 streams.
Purge
time
Typical overall time,
including time for the
BioTector to inject the
sample and flush its lines.
0
40s
76s
90s
324s, 5m24s
2
49s
85s
90s
336s, 5m36s
4
59s
95s
90s
348s, 5m48s
6
83s
119s
90s
378s, 6m18s
7
122s
158s
90s
427s, 7m07s
Sample distance, 80 meters, typical filling and purging times.
Lift,
meters
Time for venturi
sampler to fill tubes
and chamber
Venturi filling time,
including pre-purge
times for 4 streams.
Add 8s for 6 streams.
Purge
time
Typical overall time,
including time for the
BioTector to inject the
sample and flush its lines.
0
124s
160s
120s
459s, 7m39s
2
145s
181s
120s
486s, 8m06s
4
179s
215s
120s
528s, 8m48s
6
255s
291s
120s
624s, 10m24s
7
345s
381s
120s
735s, 12m15s
Notes:
•
The typical overall times above are also the minimum BioTector cycle time.
•
It is essential that the tube slopes gently back to the sample point, dips and rises in the
sample tube increases the sample transport time.
•
Any point of the sample tube must never be higher than 7 meters above the sample
point.
5.
Manual Operation:
To manually fill the sampler, press the [A] key on the PLC for > 2 seconds. The PLC latches
this input, and this simulates the operation of the BioTectors SAMPLER FILL command.
To manually empty and purge the sampler, press the [B] key on the PLC for > 2 seconds. The
PLC latches this input, and this simulates the operation of the BioTectors SAMPLER EMPTY
command.
To reset the sampler at any time, press the [ESC] key.