Technical Specifications
BioTector Venturi Sampler
Technical Specifications
TYPICAL TECHNICAL DATA
PLC Enclosure:
Polystyrene with glazed polycarbonate door
Dimensions (HxWxD):
2010mm x 390mm x 160mm
Weight:
20 kg
– 30 kg
Noise level:
50-70 dB(A) during filling and purging with short term noise 1 to 2
seconds in duration up to 90 dB(A) when pneumatic valves are
de-energized.
Power Consumption:
24 W (1 Amp 24V DC)
Control Wire Specification:
Number of Cores = 6 cores, Current Rating minimum = 3 Amp, CSA
(Cross Sectional Area minimum) = 0.5mm
2
.
FEATURES IN DETAIL
Operation:
PLC with Keyboard
Automatic no sample fill fault indication
Direct control from the BioTector through the 6 core cable
Sample chamber and sample line purge with air and water
Language Options:
English (Other languages are available upon request.)
OPTIONAL FEATURES
Multi-stream:
Valves for up to 6 streams
CONSUMABLES
Instrument Air:
6 bars
(50 liters/minute flow rate when venturi pump is running at 6bar)
(~100 liters/minute flow rate when venturi sampler is purging at 3.5bar)
Note: If using a standalone compressor please ensure it can continually
supply the above requirements, this may require up to 100% duty cycle.
Air Quality:
-20°C dew point (free of water, oil and dust)
Water:
3 bars (~10 liters per hour at 3 bars)
Water can be heated up to 90°C for hot water backwash
Water regulator and/or water heater to be supplied by the user
Service:
Typical 6 Monthly Intervals
SYSTEM PARAMETERS
Sample Chamber:
PFA
Sample Volume:
70ml available to BioTector
Fittings and Valves:
SS
-
316, EPDM, PTFE
Venturi Pump:
SS
-
316
Sampler Sample Tubing:
PFA (1/2” OD x 3/8” ID)
Cycle Time:
Controlled by BioTector