17
AV delay
AV hystereses
Ventricular pacing
Upper rate
Mode switching
Parameter
Range of values
Standard
SR DR HF
AV delay
Low; Medium; High; Fixed;
Individual
Low
x
x
20 ... (5) ... 350 ms
(in 6 rate ranges)
180-170-160-
150-140 ms
x
150-140-130-
120 ms
x
Sense compensation
OFF; -10 ... (-5) ... -120 ms
-45 ms
x
x
AV safety interval
100 ms
100 ms
x
x
Parameter
Range of values
Standard
SR DR HF
AV hysteresis mode
OFF; Positive; Negative
HF in RV modes: IRSplus
OFF
x
x
Positive modes:
AV hysteresis
70; 110; 150; 200 ms
70 ms
CLS modes:
110 ms
x
x
Negative modes:
AV hysteresis
10 ... (10) ... 150 ms
50 ms
x
x
AV repetetive / scan
cyles
ON; OFF
ON
x
x
Parameter
Range of values
Standard
SR DR HF
Ventricular pacing
BiV; RV; LV
BiV
x
Triggering
OFF; RVs; RVs + PVC
RVs
x
LV T-wave protection
ON; OFF
ON
x
Maximum trigger rate AUTO; 90 ... (10) ... 160 bpm
AUTO
x
Initially paced
chamber
RV; LV
LV
x
VV delay after Vp
0 ... (5) ... 80 ... (10) ... 100 ms 0 ms
x
VV delay after sense
0 ms
0 ms
x
Parameter
Range of values
Standard
SR DR HF
Upper rate
SR: in VVT mode
90 ... (10) ... 200 bpm
130 bpm
x
x
x
Wenckebach response
Automatically set
—
x
x
Atrial upper rate
OFF;
175; 200; 240 bpm
240 bpm
x
x
Parameter
Range of values
Standard
SR DR HF
Mode switching
OFF; ON
ON
x
x
Intervention rate
100 ... (10) ... 250 bpm
160 bpm
x
x
Switch to (mode)
DDI; DDI(R) when
permanent DDD(R)
VDI; VDI(R) when
permanent VDD(R)
DDI(R)
VDI(R)
x
x
Ventricular pacing
RV; BiV
BiV
x
Onset criterion
3 ... (1) ... 8
5
x
x
Resolution criterion
3 ... (1) ... 8
5
x
x
Change of the basic rate
with mode switching
OFF; +5 ... (5) ... +30 bpm +10 bpm
x
x
Rate stabilization with mode
switching
OFF; ON
OFF
x
x
2:1 lock-in protection
OFF; ON
DR: ON
HF: only
when RV
modes ON
x
x
Parameter
Range of values
Standard
SR DR HF
401303--G_GA_Eluna-I-ProMRI_mul.book Page 17 Friday, October 2, 2015 4:57 PM