18
Ventricular pacing suppression
Parameters valid for devices in DDD-ADI or DDDR-ADIR modes:
Refractory periods
Blanking periods
PMT protection
Pacing and Sensing
Pulse amplitude and pulse width
Sensitivity
Parameter
Range of values
Standard
SR DR HF
Vp suppression
ON; OFF
OFF
x
x
Pacing suppression after
consecutive Vs
1 ... (1) ... 8
6
x
x
Pacing supported after X-out-
of-8 cycles
1; 2; 3; 4
3
x
x
Parameter
Range of values
Standard
SR DR HF
Refractory period
200 ... (25) ... 500 ms
250 ms
x
Atrial refractory
period
AUTO
AUTO
x
x
Atrial refractory
period in the modes
AAI(R); AAT(R); DDT
300 ... (25) ... 775 ms
350 ms
x
x
AUTO PVARP
ON; OFF
ON
x
x
PVARP
Auto PVARP OFF:
175 ... (25) ... 600 ms
Auto PVARP ON:
Automatically set
x
x
PVARP after PVC
PVARP + 150 ms
(max: 600 ms) is auto-
matically programmed
400 ms
x
x
Right ventricular
refractory period
200 ... (25) ... 500 ms
250 ms
x
x
Left ventricular
refractory period
200 ms
200 ms
x
Parameter
Range of values
Standard
SR DR HF
Far-field protection
after Vs
100 ... (10) ... 220 ms
100 ms
x
x
Far-field protection
after Vp
100 ... (10) ... 220 ms
150 ms
x
x
Ventricular blanking
period after Ap
30 ... (5) ... 70 ms
30 ms
x
x
Parameter
Range of values
Standard
SR DR HF
PMT protection
OFF; ON
ON
x
x
VA criterion
250 ... (25) ... 500 ms
350 ms
x
x
Parameter
Range of values
Standard
SR DR HF
Pulse amplitude
A/RV/LV
0.2 ... (0.2) ... 6.0 ... (0.5)
... 7.5 V
3.0 V
x
x
x
Pulse width A/RV/LV
0,1 ...(0,1) ... 0.5 ... (0.25)
... 1.5 ms
0.4 ms
x
x
x
Parameter
Range of values
Standard
SR DR HF
Sensitivity
AUTO; 0.5 ... (0.5) ... 7.5 mV
AUTO
x
Sensitivity A
AUTO; 0.1 ... (0.1) ... 1.5 ... (0.5)
... 7.5 mV
AUTO
x
x
RV sensitivity
AUTO; 0.5 ... (0.5) ... 7.5 mV
AUTO
x
x
LV sensitivity
OFF; AUTO; 0.5 ... (0.5) ...
7.5 mV
AUTO
x