Atrial capture control
Ventricular capture control
Atrial overdrive pacing
Lead configuration
IEGM recordings
Parameter
Range of values
Standard
SR DR HF
Atrial capture control
ATM (monitoring only); ON;
OFF
ON
x
x
Minimum amplitude
0.5 ... (0.1) ... 4.8 V
1.0 V
x
x
Threshold test start
2.4 ... (0.6) ... 4.8 V
3.0 V
x
x
Safety margin
0.5 ... (0.1) ... 1.2 V
1.0 V
x
x
Search type
Interval; time of day
Time of day
x
x
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
24 h
x
x
Time of day
00:00 ... (10 min) ... 23:50
hh:mm
02:00 AM
hh:mm
x
x
Parameter
Range of values
Standard
SR DR HF
Ventricular capture
control RV, LV
ATM (monitoring only); ON;
OFF
ON
x
x
Minimum amplitude
0.7 V
0.7 V
x
x
Threshold test start
2.4 ... (0.6) ... 4.8 V
3.0 V
x
x
RV safety margin
0.3 ... (0.1) ... 1.2 V
0.5 V
x
x
LV safety margin
1.0; 1.2 V
1.0 V
x
Search type
Interval; time of day
Time of day
x
x
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
24 h
x
x
Time of day
00:00 ... (00:10)
... 23:50 hh:mm
00:30 hh:mm x
x
Parameter
Range of values
Standard
SR DR HF
Atrial overdrive pacing ON; OFF
With ON: maximum over-
pacing rate 120 bpm, mean
rate increase approximately
8 bpm, rate decrease after
20 cycles
OFF
x
x
Parameter
Range of values
Standard
SR DR HF
Sensing polarity A
Unipolar, Bipolar
Unipolar
x
x
Sensing polarity RV
Unipolar, Bipolar
Unipolar
x
x
x
Sensing polarity LV
Unipolar, Bipolar
Unipolar
x
Pacing polarity A
Unipolar, Bipolar
Unipolar
x
x
Pacing polarity RV
Unipolar, Bipolar
Unipolar
x
x
x
Pacing polarity LV
LV tip -> LV ring
LV tip -> RV ring
LV ring -> LV tip
LV ring -> RV ring
LV tip -> housing
LV ring -> housing
LV tip –>
housing
x
Parameter
Range of values
Standard
SR DR HF
IEGM recordings
8 series:
20 (quantity); each max. 10 s
—
x
x
x