20
Rates for statistics
Rate Adaptation
CLS modes: closed loop stimulation
Parameters valid for 8 series devices:
R modes: Accelerometer
MRI Program
MRI modes
Modes valid for devices marked ProMRI
Types of
IEGM recordings
High atrial rate (HAR)
HAR
x
x
x
Mode switching
x
x
x
High ventricular rate (HVR)
ON
x
x
x
8 series:
Patient triggering (triggered
by patient)
OFF
x
x
x
Pre-trigger recording 0; 25; 50; 75; 100%
75%
x
x
x
IEGM signal
Filtered; Unfiltered
Filtered
x
x
x
Parameter
Range of values
Standard
SR DR HF
High atrial rate:
HAR limit
100 ... (5) ... 250 bpm
200 bpm
x
x
High ventricular rate:
HVR limit
150 ... (5) ... 200 bpm
180 bpm
x
x
x
HVR counter
4; 8; 12; 16 cycles
8 cycles
x
x
x
Start resting period
00:00 ... (00:10) ...
23:50 hh:mm
02:00 hh:mm x
x
x
Duration of resting
period
00:30 ... (00:30) ... 12:00
hh:mm
04:00 hh:mm x
x
x
Enable lead check
ON; OFF
ON
x
x
x
Parameter
Range of values
Standard
SR DR HF
Maximum CLS rate
80 ... (5) ... 160 bpm
120 bpm
x
x
x
CLS response
Very low; Low; Medium; High;
Very high
Medium
x
x
x
Parameter
Range of values
Standard
SR DR HF
CLS resting rate
control
OFF; +10 ... (10) ... +50 bpm
+20 bpm
x
x
x
Vp required
Yes; No
No
BiV modes:
Yes
x
x
x
Parameter
Range of values
Standard
SR DR HF
Sensor gain
AUTO; Very low; Low;
Medium; High; Very high
AUTO
x
x
x
Max. sensor rate
80 ... (5) ... 180 bpm
120 bpm
x
x
x
Sensor threshold
Very low; Low; Medium; High;
Very high
Medium
x
x
x
Rate fading
OFF; ON
OFF
x
x
x
Rate increase
1; 2; 4; 8 bpm/cycle
2 bpm/cycle
x
x
x
Rate decrease
0.1; 0.2; 0.5; 1.0 bpm/cycle
0.5 bpm/
cycle
x
x
x
Mode
Range of values
Standard
SR DR HF
MRI Program
ON; OFF
OFF
x
x
x
MRI mode
OFF; A00; V00
—
x
OFF; D00; A00; V00
—
x
OFF; D00; A00; V00;
D00-BiV; V00-BiV
—
x
Parameter
Range of values
Standard
SR DR HF
401303--G_GA_Eluna-I-ProMRI_mul.book Page 20 Friday, October 2, 2015 4:57 PM