en • English
5
NBG codes
AAIR or VVIR is the NBG code for the antibradycardia mode of the single-chamber
device:
DDDR is the NBG code for the antibradycardia mode of the dual-chamber device:
DDDRV is the NBG code for the antibradycardia mode of the triple-chamber device:
BIOTRONIK Home Monitoring
®
In addition to effective pacing therapy, BIOTRONIK provides a complete therapy
management system:
• With Home Monitoring, diagnostic and therapeutic information and technical data
are automatically sent to a stationary or mobile transmitter via an antenna in the
device header. The data are encrypted and sent from the transmitter to the
BIOTRONIK Service Center via the cellular phone network.
• The received data are deciphered and evaluated. Each physician can set the criteria
for evaluation to be used for each patient and can configure the time of notification
via e-mail, SMS or fax.
• A clear overview of the results of this analysis is displayed for the attending physi-
cians on the protected internet platform Home Monitoring Service Center (HMSC).
• Data transmission from the device is performed with a daily device message.
• Device messages which indicate special events in the patient's heart or in the
device are forwarded immediately.
• A test message can be initiated at any time using the programmer to immediately
check the Home Monitoring function.
DR(-T)
• VVI-CLS; DDD-CLS (8 series only)
• DDDR; DDIR; DVIR; D00R
VDDR; VDIR; VVIR; V00R
AAIR; A00R
• DDD; DDT; DDI; DVI; D00
VDD; VDI; VVI; VVT; V00
AAI; AAT; A00
• DDD-ADI; DDDR-ADIR
• OFF
DDDR
HF-T
• DDD-CLS; VVI-CLS
• DDDR; DDIR; DVIR; D00R
VDDR; VDIR; VVIR; V00R
AAIR; A00R
• DDD; DDT; DDI; DVI; D00
VDD; VDI; VVI; VVT; V00
AAI; AAT; A00
• DDD-ADI; DDDR-ADIR
• OFF
DDDR
Note:
Home Monitoring is possible in all modes.
A/V
Pacing in the atrium or ventricle
A/V
Sensing in the atrium or ventricle
I
Pulse inhibition in the atrium and ventricle
R
Rate adaptation
Device type
Modes
Standard
D
Pacing in the atrium and ventricle
D
Sensing in the atrium and ventricle
D
Pulse inhibition and pulse triggering
R
Rate adaptation
D
Pacing in the atrium and ventricle
D
Sensing in the atrium and ventricle
D
Pulse inhibition and pulse triggering
R
Rate adaptation
V
Multisite pacing in both ventricles
401303--G_GA_Eluna-I-ProMRI_mul.book Page 5 Friday, October 2, 2015 4:57 PM