en • English
15
Timing: Triple-Chamber
Basic rate day/night
Rate hysteresis
AV delay, VV delay
AV hystereses
Ventricular pacing suppression
Upper rate
Mode switching
Parameter
Range of values
Standard
Basic rate
30 ... (1) ... 88 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 bpm
60 bpm
Night rate
OFF
30 ... (1) ... 88 ... (2) ... 122 ... (3) ... 140
... (5) ... 200 bpm
OFF
Night begins
00:00 ... (10 min) ... 11:50 PM hh:mm
10:00 PM hh:mm
Night ends
00:00 ... (10 min) ... 11:50 PM hh:mm
6:00 AM hh:mm
Parameter
Range of values
Standard
Rate hysteresis
OFF
-5 ... (-5) ... -90 bpm
OFF
Repetitive hysteresis
OFF; 1 ... (1) ... 15
5
Scan hysteresis
OFF; 1 ... (1) ... 15
5
Parameter
Range of values
Standard
AV delay
Low; medium; high; fixed; individual
Low
15 ... (5) ... 350 ms
(in 6 rate ranges)
150 to 120 ms
Sense compensation
OFF
-10 ... (5) ... -120 ms
-45 ms
AV safety interval
100 ms
100 ms
VV delay after Vp
0 ... (5) ... 80; 90; 100 ms
0
Parameter
Range of values
Standard
AV hysteresis
OFF
Negative, low; medium; high; IRSplus
OFF
All modes except for
negative ones:
AV repetitive hysteresis
OFF
1 ... (1) ... 10
OFF
Negative modes:
AV repetitive hysteresis
OFF
1 ... (1) ... 10 ... (5) ... 100 ... (10) ... 180
OFF
Parameter
Range of values
Standard
Vp suppression
OFF; ON
OFF
Pacing suppression after consecutive
ventricular sensing
1 ... (1) ... 8
6
Pacing supported after X-out-of-8 cycles
1; 2; 3; 4
3
Parameter
Range of values
Standard
Upper rate
90 ... (10) ... 200 bpm
130 bpm
Upper rate, atrium
OFF
240 bpm
240 bpm
Parameter
Range of values
Standard
Mode switching
OFF; ON
ON
Intervention rate
100 ... (10) ... 250 bpm
160 bpm
Switch to (mode)
DDI; DDI(R) when permanent DDD(R)
VDI; VDI(R) when permanent VDD(R)
DDI(R)
VDI(R)
Onset criterion
3 ... (1) ... 8
5
Resolution criterion
3 ... (1) ... 8
5
387513--D_GA_Evia-HF-ProMRI_mul.book Page 15 Thursday, September 12, 2013 3:34 PM