Chapter 5
Programmable Parameters
Evia HF / HF-T Technical Manual
PAGE 27
+/-
Minimum
Threshold
0.5 mV
Detection Hold-off Period
(DHP)
2x Step
Duratio
n
+/- Sensitivity
Threshold
from the previous event
Event
Detection
Pace
event
Atrium
DHP 120 ms
Step Duration 82 ms
Post Pace Event – Atrium -
Unipolar
Fixed value
Atrium – 1.0
mV
½
Fixed
value
Figure 13: Automatic Sensitivity Control for Atrial Unipolar Paced Events
Ventricular Pacing Overview
shows a ventricular pacing overview and applies to bradycardia programming and Mode Switch
programming.
Parameter
Range
Default
Ventricular pacing
BiV, RV
BiV
Triggering
OFF, RVs, RVs+PVC
RVs
LV T-wave protection
ON, OFF
ON
Maximum trigger rate Auto, 90...(10)...160 ppm
Auto
Initially paced
chamber
LV, RV
LV
VV Delay after Vp
0...(5)...100 ms
0 ms
VV Delay after Vs
0 ms
0 ms
Table 5: Ventricular Pacing Parameters
Figure 14: Ventricular Pacing Parameter Screen
Summary of Contents for Evia HF
Page 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Page 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Page 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Page 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Page 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Page 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Page 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Page 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Page 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Page 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...