Chapter 10
Follow-up Procedures
Evia HF / HF-T Technical Manual
PAGE 103
10.4.4.7 Sensitivity
Atrium: Auto, 0.1 - 7.5 mV. Default of Auto
Ventricle: Auto, 0.5 - 7.5 mV Default of Auto
Independently programmable for each chamber.
10.4.4.8 Pacing Polarity
Unipolar, Bipolar. Independently programmable for each chamber.
10.4.4.9 Sensing Polarity
Unipolar, Bipolar. Independently programmable for each chamber.
Especially with unipolar sensing functions, the selected sensitivity level should be checked for possible
interference from skeletal myopotentials. If oversensing is observed, the programming of a lower
sensitivity (higher value), or bipolar sensing function, if the implanted lead is bipolar, should be evaluated.
10.5 Testing for Retrograde Conduction
The Retrograde conduction test is performed in a VDI mode with a programmable ventricular
pacing rate.
The Evia HF / HF-T allows the user to choose RV or LV pacing for the Retrograde Conduction test.
Select Test mode to display the options (
Figure 63
).
Figure 63: Retrograde Test Mode Options
Figure 64: Retrograde Conduction Test with a Negative Result
Summary of Contents for Evia HF
Page 1: ...Cardiac Rhythm Management Heart Failure Therapy Evia HF HF T Evia HF HF T Technical Manual...
Page 8: ...Chapter Table of Contents Evia HF HF T Technical Manual PAGE vi...
Page 12: ...Chapter 2 Indications Evia HF HF T Technical Manual PAGE 4...
Page 14: ...Chapter 3 Contraindications Evia HF HF T Technical Manual PAGE 6...
Page 98: ...Chapter 7 Product Storage and Handling Evia HF HF T Technical Manual PAGE 90...
Page 104: ...Chapter 9 Lead Connection Evia HF HF T Technical Manual PAGE 96...
Page 118: ...Chapter 10 Elective Replacement Indication ERI Evia HF HF T Technical Manual PAGE 110...
Page 130: ...Chapter 13 Order Information Evia HF HF T Technical Manual PAGE 122...
Page 132: ...Chapter 14 Order Information Evia HF HF T Technical Manual PAGE 124...
Page 136: ...PAGE 128 Appendix A Evia HF HF T Technical Manual...