background image

Note: 

Tissue damage due to excessive heating usually causes change or loss of the 

sensing and pacing function of the implanted lead.

9

Safety

Risky procedures 

The table below provides an overview of procedures that present a risk to patients 
with an implanted lead or a device system. Take appropriate precautionary 
measures and observe the specific instructions listed in the table. 

Problematic environmental 

influences

Increased ambient pressure:

!

!

CAUTION

Damage and failure of the device system

Patients with device systems must avoid situations or environments in which they 
would be exposed to high ambient pressures (such as diving or pressure cham-
bers).

The leads are manufactured under standard atmospheric pressure and are not 
designed to withstand increased ambient pressure.
Stress resulting from excess pressure may damage the leads.

Electromagnetic interference:

Electromagnetic fields may negatively affect patients with device systems as the 
intensity and duration of exposure increase. This can have the following 
consequences:
— Temporary or permanent effect on or damage to the device system
— Induction of tachycardias, up to and including ventricular fibrillation (in rare 

cases)

— Thermal tissue damage (in severe cases)
The patient should be properly informed and instructed on behaviors to avoid 
situations with especially risky electromagnetic effects. 
Perform a follow-up for clarification if electromagnetic interference is 
suspected to have impaired the function of the device system.
In most cases, the problem can be solved by reprogramming the device.

Procedure

Type of danger

Recommendations for risk 
mitigation

Therapeutic ultra-
sound

Tissue damage due to 
excessive heating of the 
lead

Do not direct the energy focus 
onto the lead or the device.
Afterwards: perfom a full 
follow-up.

External defibrilla-
tion

Tissue damage due to 
excessive heating of the 
lead

Afterwards: perfom a full 
follow-up.

Electrophysiological 
ablation

Tissue damage due to 
excessive heating of the 
lead
Induction of ventricular 
fibrillation
Damage to the lead

Switch off the active device 
beforehand.
Keep as much distance as 
possible between the ablator 
and the lead.
Following ablation and prior to 
restarting the active device: 
perfom a full follow-up.

HF surgery (electro-
cautery)

Tissue damage due to 
excessive heating of the 
lead,
Induction of ventricular 
fibrillation

Do not direct the energy focus 
onto the lead or the device.
Afterwards: perfom a full 
follow-up.

Lithotripsy

Mechanical effect on or 
damage to the lead

Keep energy focus from the 
lead.
Afterwards: perfom a full 
follow-up.

Summary of Contents for ProMRI Protego DF-1 S

Page 1: ...les biotronik com www biotronik com Protego DF 1 S ProMRI Tripolar ICD lead with active fixation Technical Manual 414086 Revision E 2015 05 28 BIOTRONIK SE Co KG All rights reserved Specifications sub...

Page 2: ...lectrical and Electromagnetic Safety 10 Handling and Implantation 11 Implantation Basic Instructions and Safety Measures 11 Information on the Steroid Collar 12 Opening the Package 13 Checking the Fun...

Page 3: ...risks and possible complications Note Keep this technical manual for later use This technical manual This technical manual is either included in hard copy form in the product packaging or can be down...

Page 4: ...improve the lead s gliding properties Details Technical Data p 24 ff Electrical properties The fixation screw is electrically active and forms the distal pole tip electrode of the lead Sensing and pa...

Page 5: ...tines is not possible Contraindications Implantation of this lead is contraindicated in the following cases Patients with mechanical tricuspid valve prostheses or severe tricuspid valve diseases Patie...

Page 6: ...bolisms and damage to the device Resterilization and reuse are prohibited CAUTION Improper storage If the specified time period and temperature range for storage are exceeded then the documented prope...

Page 7: ...nctions of the device system which consists of pacemaker or ICD and leads Incorrect lead implantation Lead dislodgement Lead fracture Insulation defect Battery depletion or component failure of the ac...

Page 8: ...ducts are actually certified MR conditional in your country or region Procedure Type of danger Diathermy Tissue damage due to excessive heating of the lead Induction of ventricular fibrillation Magnet...

Page 9: ...s Thermal tissue damage in severe cases The patient should be properly informed and instructed on behaviors to avoid situations with especially risky electromagnetic effects Perform a follow up for cl...

Page 10: ...electromagnetic energy as an antenna would and cause electrical voltages at the lead tip and connector This can induce ventricular fibrillation in some cases as well as damage or otherwise affect the...

Page 11: ...of the stylet in the lead and inhibit or block the screw mechanism Ensure that no blood reaches the interior of the lead on or with the stylet As far as possible prevent blood from entering the lead...

Page 12: ...ate The intended effect is the reduction of the inflammatory processes after implantation and the inflammation related post operative threshold increase lead maturation behavior Long term performance...

Page 13: ...hin the other and sterilized with ethylene oxide As a result the inner blister is also sterile on the outside You can remove the inner blister by using a standard aseptic technique and place it in the...

Page 14: ...chanism are not suitable for implantation Do not implant the lead if it fails the function test Instead use a replacement lead that has passed the same test Step Action 1 Remove the stylet guide from...

Page 15: ...sing loss due to excessive mechanical stress of the lead Make sure the lead does not become pinched between the clavicle and the first rib after implantation After having established the access to the...

Page 16: ...crew is retracted Thus the position of the lead tip can be assessed prior to fixation using electrocardiac measurements without injuring the tissue Fixating the lead tip Proceed as follows to fixate t...

Page 17: ...ully when fixating the lead tip Consider the following two aspects when elongating between the distal and proximal fixations of the lead The patient s own movement and heart contractions should not ex...

Page 18: ...asurements pacing will be temporarily interrupted Connecting the leads temporarily For a qualitative evaluation of the lead position it is necessary to measure pacing thresholds intracardiac potential...

Page 19: ...the respective testing and measuring devices for further details on performing measurements and tests Measuring the threshold In order to measure the pacing threshold the pacing rate of the measuring...

Page 20: ...incision point and decreases the risk of damaging the insulation or coil during fixation Prerequisites Placement of the lead and measurement of the threshold and the intracardiac signals were success...

Page 21: ...out damaging the thread Note Ensure that connections are clean Note Further information on connecting the lead to the active device s IS 1 and DF 1 ports can be found in the technical manual of the de...

Page 22: ...l the connector port of the active device is completely clear 4 Push the lead connector into the port without bending the conductor until the connector tip becomes visible behind the set screw block P...

Page 23: ...ocal pressure and abrasion can damage the lead insulation Depending on the implantation site and patient s anatomy the lead may be longer than required to connect the active device and position the le...

Page 24: ...In the ventricle bipolar Application Fixation in the right ventricle ventricular sensing pacing and defibrillation Shock coil RV Connections 1 x IS 1 1 x DF 1 Diameter 2 6 mm 7 8 F Insulation Silicone...

Page 25: ...m2 Material Platinum iridium alloy Surface structure Iridium fractal Distance to the lead tip 11 mm Length 50 mm Diameter 2 6 mm Area 290 mm2 Material Platinum iridium alloy Distance to the lead tip 1...

Page 26: ...a registered trademark for a particular cobalt chromium nickel alloy Design Cable DFT wires per cable 7 x 7 Material of conductor MP35N DFT Material of insulation Teflon PFA Outer diameter 0 28 mm inc...

Page 27: ...vents will not occur Lead malfunctions or failures Defense reactions of the body against lead implantation Medical complications including myocardial perforation during lead implantation or as a conse...

Page 28: ...packaging is damaged CE mark Unipolar IS 1 connector Bipolar IS 1 connector Unipolar DF 1 connector DF4 connector for ICD leads with one shock coil DF4 connector for ICD leads with two shock coils IS...

Page 29: ...ssible guide wire diameter Total length Surface and material of the indicated lead Recommended size of the lead introducer Additional stylets as part of the package contents Lead fixation sleeve premo...

Page 30: ...ppendix LV Positioning the lead in the coronary venous system for left ventricular pacing CS Coronary sinus Pace Pacing Sense Sensing Shock Shock DXA Dexamethasone acetate as steroid eluant Symbol Mea...

Reviews: