Calibration
17
Chapter 2
Calibration
Calibration Accessories
y
Calibration Combination (Cal Combo) which
contains:
One 50 ohm load
One Open standard
One Short standard
y
Test Cable (optional) – A phase-stable cable will
provide consistent, reliable results.
Note:
When using a test cable connected to the
Site Analyzer, attach the Cal Combo to the end of the
cable during calibration.
Calibrating
For best results, set the frequency and calibrate the
Bird Site Analyzer immediately before taking mea-
surements.
Follow the numbered steps in Figure 2, page 18 to cal-
ibrate the Site Analyzer.
Note:
When calibrated, the Site Analyzer’s main
screen will display “Calibration: FULL”. When uncali-
brated, the Site Analyzer’s main screen will display
“Calibration: OFF”.
Note:
Calibration is performed at the highest
number of data points (949 points) and the calibration
graphs are displayed using the currently selected dis-
play method (envelope, maximum, or minimum).