Measure Match Mode
33
Figure 10 Setting a Limit Line
800.00
MHz
1600.00
Calibration: Full
Limit: 3.00
FAIL
Save/
Recall
Auto
Scale
Limit
Test ON
HOLD
Measure Match
[
]
238 Pts
10.00
1.00
V
S
W
R
(0.00
/ )
2
1
4
5
3
Step
Action
1
While in the Measure Match mode or the
Fault Location mode, press the Limit Test
key:
First press - Test ON Aud (line and sound)
Second press - Test ON (line only)
Third press - Test OFF
2
Press the Config key
3
Press the Limit Line soft key
4
Use the keyboard to enter a limit value of
your choice or use the suggested limits
listed below (use the +/- key to enter a neg-
ative value) then press the Enter key.
Antenna at resonance
-14 dB
Connectors
-25 dB
Jumpers
-35 dB
Lightning protector
-25 dB
Transmission lines
-30 dB
5
Press the Esc key to exit the configure
screen