Bird Site Analyzer
136
Troubleshooting
Any service procedure not covered in this manual
should be referred to an authorized service facility.
Locate the problem, review the possible causes, and
perform the action listed. If the problem is not cor-
rected, give us a call or return the unit for service.
Problem
Possible Cause
Possible Correction
Unit will not power up Battery pack drained
Charge the battery pack.
Battery pack unable to
keep a charge
Replace the battery pack.
AC adapter is damaged Replace the ac adapter.
Self test fails
Error condition
Turn the unit off and then
back on. If the problem
persists, return the unit
for service.
Date and time appear,
hiding the display
Internal error
Press
Mode
,
Utility
,
and then
Escape
.
Fault Location trace
appears incorrect
Scale too large
Press
Auto Scale
to
make the scale smaller.
Incorrect cable loss or
velocity of propagation
Check the cable loss and
V
p
settings.
Erratic antenna test
measurements
Bad calibration combo
Use a different calibration
combo.
Soft keys beep when
pressed without
performing any
function
Context switching
operations taking place
Wait for screen changes to
complete before.
Keys do not respond
Unit is “Locked Up”
Turn the unit off and then
back on.
Perform a full system
initialization. Disconnect
the AC adapter, hold down
the Contrast key and turn
the unit off. Turn the unit
back on.
Summary of Contents for SITE ANALYZER SA-1700EX
Page 2: ...This page is not blank...
Page 8: ...Bird Site Analyzer vi...
Page 9: ...vii...
Page 30: ...Bird Site Analyzer 16...
Page 142: ...Bird Site Analyzer 128...
Page 148: ...Bird Site Analyzer 134...
Page 159: ...Maintenance 145 Measured Return Loss dB...
Page 167: ...Maintenance 153 the number of data points you wish measure 238 475 949...