Bird Site Analyzer
18
Figure 2 Calibrating the Site Analyzer
8
800.00
MHz
1600.00
Calibration
Calibration: OFF
Load
Short
Open
10.00
1.00
V
S
W
R
(
0.00
/
)
6
7
1
5
2
3
4
Step
Action
1
Press the Calibrate key
2
Connect the Cal Combo “Open” connector to
the unit antenna port
3
Press the Open soft key then wait for the
unit to beep and display Done
4
Connect the Cal Combo “Short” connector to
the unit antenna port
5
Press the Short soft key then wait for the
unit to beep and display Done
6
Connect the Cal Combo “Load” connector to
the unit antenna port
7
Press the Load soft key then wait for the
unit to beep and display Done
8
Remove the Cal Combo from the unit
Summary of Contents for SITE ANALYZER SA-1700EX
Page 2: ...This page is not blank...
Page 8: ...Bird Site Analyzer vi...
Page 9: ...vii...
Page 30: ...Bird Site Analyzer 16...
Page 142: ...Bird Site Analyzer 128...
Page 148: ...Bird Site Analyzer 134...
Page 159: ...Maintenance 145 Measured Return Loss dB...
Page 167: ...Maintenance 153 the number of data points you wish measure 238 475 949...