Bird Site Analyzer
16
y
Press the
CALIBRATE
key.
y
Attach the open standard.
Press the
OPEN
key.
Wait for a “beep” and for the trace to scroll before
continuing.
y
Attach the short standard.
Press the
SHORT
key.
Wait for a “beep” and for the trace to scroll before
continuing.
y
Attach the load standard.
Press the
LOAD
key.
Wait for a “beep” and for the trace to scroll before
continuing.
y
Press the
CAL DONE
key. New coefficients are
calculated. The Site Analyzer is now calibrated.
☞
NOTE: Now is a good time to save the setup.
Saving after calibration will save the new
calibration coefficients. See “Save Setup”, page 75.