15
Chapter 2
Calibration
Calibrating the SiteHawk Analyzer
For best results, set the frequency and calibrate the Bird Site Analyzer
immediately before taking measurements.
Note:
Use a precision open, short, load (OSL) calibration standard or
“Calibration Combo” for accurate measurement results.
Note:
When using an extension cable, a phase stable cable is needed
to ensure performance and accuracy.
Note:
Changing frequency settings will automatically turn calibration
off. Always set the frequency before calibrating the unit.
1.
Press the Calibrate icon.
2.
Connect the Cal Combo “Open” connector to the unit antenna port.
3.
Press the Open softkey.
4.
Wait for the unit to sweep and a Green Check to display.
5.
Connect the Cal Combo “Short” connector to the unit antenna port.
6.
Press the Short softkey
7.
Wait for the unit to sweep and a Green Check to display.
8.
Connect the Cal Combo “Load” connector to the unit antenna port.
9.
Press the Load softkey
10. Wait for the unit to sweep and a Green Check to display.
Note:
During calibration, the color of the icon display is yellow.
Note:
After calibration, the display will resemble
and Correction will be set to On.
Figure
7
Calibration Display Screen