8
Chapter 5
Maintenance
This chapter covers cleaning, inspection, troubleshooting, and specifications for the Bird 8251 series loads.
Troubleshooting
The table below contains troubleshooting information for problems which can occur during normal operation. This
manual cannot list all malfunctions that may occur, or their corrective actions. If a problem is not listed or is not
corrected by the listed actions, notify a qualified service center.
Maintenance
Cleaning
The outside surface of the instrument should be wiped free of dust and dirt when necessary. Excessive dust on the
cooling fins will interfere with heat dissipation. Clean the RF connector, both metallic and insulating surfaces, with a
dry, non-residue forming solvent.
Problem
Possible Cause
Correction
Leaking coolant
Loose clamping band
Tighten the clamping band
Defective or improperly installed front
O-ring
Replace the O-ring
(
Defective or improperly installed rear
O-ring
Replace the O-ring
(
High or low DC resistance
Loose RF input connector
Tighten connector
Faulty RF input connector
Models with “QC” type connectors: Replace connector
(
All other models: Return the unit for service
Faulty resistor
Replace the resistor
(
Overheating radiator
RF power too high
Lower RF power (
for maximum RF power)
Coolant level too low
Check the coolant level. Add coolant if necessary (
Coolant degraded
Replace the coolant
(
)
Faulty resistor
Replace the resistor
(
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line while RF
power is being applied. Leaking RF energy is a potential health hazard.
WARNING
Disconnect the unit from all power sources before servicing. The unit may be energized
from multiple sources. The potential for electric shock exists.