background image

BT-410F and BT-410A Operation manual

 

P/N:410-ENG-OPM-EUR-R12 

Bistos Co.,Ltd. 

2021,05 

12 

5 Troubleshooting

 

Problems 

Possible cause 

Corrective action 

Headlight does not turn on. 

Battery  is  discharged  or  is 
defective. 

Recharge the battery pack. 
If the device does not turned 
on after charging, contact to 
the service center 

Brightness  decreases  during 
use. 

Temperature of LED may have 
increased due to blockage of 
airflow. 

Turn off the headlight, check 
and remove any obstructions 
from the headlight module and 
vent hole. 

Brightness is diminished 
and/or the spot appears fuzzy. 

Headlight lens may be 
contaminated. 

Clean the lens according to this 
manual. 

Fully charged battery operates 
for less than 2 hours. 

Battery may be losing its 
capacity. 

Contact to the service center 
to replace battery pack. 

  If you have not resolved the problem with device or you have further questions, please contact Bistos 

customer center.

 

 

6 Specifications

 

The BT-410F and BT-410A head-worn light is classified as follows in accordance with EN/IEC 60601-1 

 

Category Name 

Specification 

Type of electric shock protection 

Internally powered equipment 

Operating mode 

Continuous mode 

 

Main body   

Dimensions 

303(L) X 165(H) X 195(D) mm 

Illumination 
(at 25 cm working distance) 

15,000 ~ 30,000 lx 
Maximum illumination with Astral lamp: 50,000 lx (Applicable 
model: BT-410A) 

LED lift time 

50,000 hours 

Headband size 

534 ~ 638 mm 

Weight 

119 g 

 

Battery pack 

Type 

Lithium-ion rechargeable battery 

Capacity 

3.7V, 2,600mAh 

Continuous operating time 

4 hours 

Life span 

800 charge-discharge cycles 

Weight 

69 g 

 

Battery charger 

Input 

AC 100 ~ 240 Vac, 50/60 Hz 

Output 

DC 4.2V, 1000 mA 

Charging time 

4 hours (100%) 

 

Environment 

Operating 

Temperature 

10°C to 40°C (50°F to 104°F) 

Humidity 

5 ~ 85% non-condensing 

Atmospheric pressure 

70 -106kPa 

Transport and storage 

Temperature 

- 20°C to 60°C (–4°F to 140°F) 

Humidity 

0% to 95% non-condensing 

Summary of Contents for BT-410A

Page 1: ...BT 410F and BT 410A Head worn light Operation Manual Keep this manual for future reference P N 410 ENG OPM EUR R12...

Page 2: ...e However no responsibility is assumed by Bistos Corporation For its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implicat...

Page 3: ...7 2 1 Product outlook and description 7 2 2 Loupe optional assembly 8 2 3 Astral lamp optional assemble 9 3 Operation 9 3 1 Adjust the headband 9 3 2 Turn on the power 9 3 3 Focusing the light 9 3 4...

Page 4: ...n product label packaging and this manual in order to stand for the information about Stand by Those symbols identify safety information Warning and caution necessarily well know when operating the de...

Page 5: ...old environment to a warm humid location Do not attempt to service the BT 410F and BT 410A Only qualified service personnel should attempt any needed internal servicing 0 3 Battery warnings WARNING Do...

Page 6: ...rief Device Description Product Features Model Configurations 1 2 Brief device description This Head worn light is used for medical examination by fixing it on the head and illuminating with condensed...

Page 7: ...0A has Fixed Focus lens but BT 410F has Variable focus lens BT 410F and BT 410A system consists of the following function and elements Model BT 410F BT 410A Control of Light Intensity X O Control of L...

Page 8: ...ration 2 1 Product outlook and description Fig 2 1 BT 410A outlook and description Headband height adjustable part Headband part Headband size adjustable part Protect band Power On Off button LED indi...

Page 9: ...e shown below Fig 2 3 Loupe Assembly 2 3 Astral lamp optional BT 410A only assembly The astral lamp can be assembled at the bottom of light source Remove the lens cap from the light source and connect...

Page 10: ...e high charging voltage may cause the device malfunction 3 3 Focusing the light BT 410F only You can adjust the focal length between 25 30 cm Adjust the focus for the best light quality before treatme...

Page 11: ...r is unseparated When charging unseparated be sure the device turned off To separate or assemble the battery pack refer below Fig 3 1 Battery pack assembly method Fig 3 2 Battery pack separation metho...

Page 12: ...s during surgery or other cases 4 Maintenance and cleaning 4 1 Checking the light intensity It is recommended that the intensity of the light be checked before each use and at least every six months W...

Page 13: ...h device or you have further questions please contact Bistos customer center 6 Specifications The BT 410F and BT 410A head worn light is classified as follows in accordance with EN IEC 60601 1 Categor...

Page 14: ...ons flicker emissions IEC61000 3 3 Complies WARNING The BT 410F and BT 410A should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the device should be...

Page 15: ...equency magnetic fields should be at levels characteristic of a typical commercial or hospital environment Conducted RF IEC61000 4 6 3 Vrms 150 kHz to 80 MHz 6 Vrms in ISM bands between 0 15 MHz and 8...

Page 16: ...d with the following symbol IEC 60417 Ref no 5140 Non ionizing electromagnetic radiation Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all si...

Page 17: ...t is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on Consumers noticed by...

Reviews:

Related manuals for BT-410A