Troubleshooting
Operating instructions iS10 / iS20
100
38019417000 en
11
Fault
Cause
Elimination
"Er22695"
The weighing range and the set
interval exceed the permissible
number of verification scale in-
tervals.
Example:
Max. 6 kg e = 0,001 kg equals
6000 verification scale intervals.
Max. 10000 verification scale in-
tervals for analog load cell.
Max. 7500 verification scale in-
tervals for digital weighing sys-
tem.
"Er22696"
Authentication of digital load cell
failed.
Switch device off and on.
"Er22697"
The data in the metrologically
approved data storage could not
be read.
Switch device off and on.
"Err Cnt"
Minimum weight for reference
weight determination is not
reached.
Place higher weight or more
pieces on the scale.
"Err toL"
Tolerance control has not been
activated or invalid parameter.
Activate tolerance control in step
50. Target value, upper and low-
er tolerance limit.