( F ) UNIT NOT TAKE CURRENT.
UNIT NOT TAKE
CURRENT
Is Current Set Preset R21 OK ?
Replace It.
Is Current Pot Short ?
Replace It.
Is Wiring Related to
Ensure proper Continuity.
Current Pot OK ?
Are Load Terminals connected
properly with Nut inside Unit ?
Ensure proper Terminal Contact.
1.
Check Current set Preset R21. It might be open.
2.
No Contact of Load Terminals with the Terminal Nut inside the Unit.
3.
Current Pot might be Shorted.
6.
Check Wiring related to Current Pot.