The 16 digit data stream will be displayed in the following format:
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1D0
Each digit indicates the following status:
D15
Start Word
D14
4
D13
Function: T1, T2, T3, T4
When sending T1 value, D13 = 1
When sending T2 value, D13 = 2
When sending T3 value, D13 = 3
When sending T4 value, D13 = 4
Function: T1, T2, T1-T2
When sending T1 value, D13 = 1
When sending T2 value, D13 = 2
When sending T1-T2 value, D13 = 3
D12 & D11
Indicator for Display
°C = 01
°F = 02
D10
Polarity: 0 = Positive / 1 = Negative
D9
Decimal Point (DP), position from right to left
0 = No DP, 1 = 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1
Display reading, D1 = LSD, D8 = MSD
Example: If the display reading is 1234,
then D8 to D1 is: 00001234
D0
End Word
RS232 Setting
Baud Rate
9600
Parity
No Parity
Data bit #
8 Data bits
Stop bit
1 Stop bit
11
Summary of Contents for Anaheim Scientific H240
Page 1: ...INSTRUCTION MANUAL Models H240 Type K J Four Channel Thermometer with Pt 100 ohm inputs ...
Page 2: ......
Page 15: ......