101
5.3
OVP Calibration
Follow the step by step instructions below for OVP calibration.
1.
Press
in the calibration menu to enter the OVP calibration
menu.
2.
Press
to access OVP calibration procedure for “Lo”.
3.
When complete, go to the OVP “Hi” parameter and press
.
4.
Once completed, the calibration values will be stored into the
internal flash memory and return to the calibration menu. If
calibration is not completed within 10 seconds after calibration
starts, there may be some hardware issues with the OVP
protection circuit and may need to be inspected by a trained
technician.