Well Numbering Convention
Facing the case from the Customer side, Well 1 is on the far right-hand end of the case. The well number
increase sequentially from right to left.
Troubleshooting
Refer to the table below for troubleshooting information.
Disconnect the case power supply by turning OFF the circuit breaker in the power supply service panel
before performing any diagnostic testing.
Problem
Cause
Possible Solution
Case Not Operating
(Clear pilot light not illuminated)
No Power to the case.
Check circuit breaker or fuses at
building power panel.
Case Not Operating
(Clear pilot light illuminated)
Power switch is off.
Reset the power switch.
Control Safety Interlock
circuit has shut case off
because one of the well
temperature probes reached
a preset value.
Failed temperature probe,
temperature probes not
connected to control in proper
sequence, well heating elements
not connected to control in proper
sequence.
Holding Temperature Not
Adequate
Case and food pans have not
been preheated.
Refer to the preheating section of
Operators Manual.
Product is below 160
°
F when
loaded into case.
Check product temperature
before loading the case.
One or more heating
elements not operating
properly.
Troubleshoot as described below.
Individual Heating Element Not
Heating
Loose wire or bad
connection.
Identify and repair wire
connection.
Failed Heating Element.
Replace failed Heating Element.
Failed Output on Control
Board.
Replace entire Control Board.
Control displays PL# and beeps
every second (# specifies
number problem well).
Probe low, shorted.
Measured temperature
< 40°F.
Replace failed temperature probe
for specified well.
Control displays PH# and beeps
every second (# specifies
number problem well).
Probe high, open. Measured
temperature
> 550°F.
Replace failed temperature probe
for specified well.
PL and PH messages can appear shortly after case is turned ON.
Control displays TL# and beeps
every second (# specifies
number problem well).
Temperature low, well probe
temperature more than 25°F
below set point.
Loose wire, failed heating element,
failed controller.
Control displays TH# and beeps
every second (# specifies
number problem well).
Temperature high, well probe
temperature more than 40°F
above set point.
TL and TH messages will not appear until after the 20 minute stabilization period.
Summary of Contents for WDCG Series
Page 1: ...Series WDCG Operation Manual Hot Case Series ...
Page 22: ...20 Wiring Diagrams ...
Page 23: ...21 ...
Page 24: ...22 ...
Page 25: ...23 ...
Page 26: ...24 ...
Page 27: ...25 ...
Page 28: ...26 ...
Page 29: ...27 ...
Page 30: ...28 ...
Page 31: ...29 ...
Page 32: ...30 ...
Page 33: ...31 ...
Page 34: ...32 ...
Page 35: ...33 ...
Page 36: ...34 ...