Additional Important Safety Information

Following Medical Device Alerts on other manufacturers’ slings, it is important that users of Steiss slings 

always follow the instructions in the user manuals provided with each sling and also the labels which are 

stitched into each sling.

In particular, it is critical to the safety of the patient that the sling is checked each and every time prior to use. 

Failure to do so may result in patient injury.

If any of the following faults are identified, the sling should not be used and immediately destroyed:

•  There are signs of general wear and tear
• 

There is loose or broken stitching on any part of the sling (including the straps)

•  There are worn areas on the attachment points
•  There are discoloured areas
•  There are areas of frayed or torn material
•  There are areas of bleached material
•  The label is illegible

Inspection is required on all areas of the sling. The following information provides some guidance on key 

areas for inspection:

Example of a sling:

Key areas for inspection include:

1. 

Sling strap attachment point

2. 

Seams

3. 

Handle attachment points

4. 

End loops on sling straps

1

1

1

1

2

2

3

4

Summary of Contents for Steiss Pivot

Page 1: ...SteissTM Pivot with Clips Sling USER MANUAL...

Page 2: ...ing in the sling and procedure for transfer has been correctly determined for the patient For further guidance please contact Blueleaf Ltd FOR USE WITH PASSIVE LIFTS ONLY Expected Service Life The exp...

Page 3: ...rough the modesty strap located on the opposite leg Position the rear section of the sling up and around the shoulders and head 1 Fitting the Sling from a Lying Position Safely position the client ont...

Page 4: ...reverse this procedure SECURI3 CLIP SYSTEM Pivot Handle Fitting the Sling to the 4 Point Cradle WARNING SECURI3 CLIP SAFETY Ensure all green safety loops are visible before lifting Once the sling is...

Page 5: ...Machine wash at 85 C DO NOT wash with bleach Bleach will damage the sling s material and make it unsafe for use Cool tumble dry air dry or dry at very low temperature DO NOT dry clean 85 C WARNING Sli...

Page 6: ...he following faults are identified the sling should not be used and immediately destroyed There are signs of general wear and tear There is loose or broken stitching on any part of the sling including...

Page 7: ...g body should be secure Stitching is cross stitched and not to be frayed Cross Stitching Typical seam stitching Photograph shows example of correct stitching in place and not frayed Typical stitching...

Page 8: ...erns Healthcare Ltd Worcestershire WR10 2AG Blueleaf Ltd Charwood House Wilberforce Way Oakhurst Business Park Southwater Horsham West Sussex RH13 9RT United Kingdom Tel 0330 055 2288 info blueleafcar...

Reviews: