Mod.BM43
COM
Vers.1.0
16/11/2005
Page 50 of
110
The following display refers to the last saved setting. The blinking digit is hiding the
number of the setting to be saved (from 0 to 17).
P a r a m e t e r s S e t u p
: : 4 0 9 . 3 4 7 m m , C a r b o n
➢
Digit
MENU
2
7
and the device will display
S a v e / L o a d p a r a m e t e r s
1 2 3 3 7 9 m m 2
This window shows the current parameters.
➢
Digit
MENU
2
8
and the device will display
H o l d o n p o o r s i g n a l
Y E S
This window is used to set up the device in order to maintain the last reading as valid in
case of temporary loss of signal. If NO is selected, output signal could be modified.
➢
Digit
MENU
2
9
and the device will display
E m p t y P i p e S e t u p
[ 2 9
4 0
This window is used to set up a min threshold. Below that threshold, the device considers
the pipe as empty. Set up a value among 30 and 40 in order to be sure the device does
not measure when the pipe is empty.
3.10
FLOWRATE UNITS MENU ANALYSIS
➢
Digit
MENU
3
0
and the it will display
M e a s u r e m e n t U n i t s
I n
0 .
M e t r i c
This window is used to set up the measuring system:
0.
metric
1.
English
Press
ENT
and then use the arrow keys to choose the measuring system. Confirm with
ENT
.
➢
Digit
MENU
3
1
and the it will display
Summary of Contents for TTFM100-F18-NG
Page 23: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 23 of 110 Or Or ...
Page 30: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 30 of 110 V mounting method N mounting method ...
Page 78: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 78 of 110 ...
Page 79: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 79 of 110 ...
Page 80: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 80 of 110 5 2 SOUND SPEEDS IN FLUIDS ...
Page 81: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 81 of 110 ...
Page 82: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 82 of 110 ...
Page 83: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 83 of 110 ...
Page 84: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 84 of 110 ...
Page 85: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 85 of 110 ...
Page 86: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 86 of 110 ...
Page 87: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 87 of 110 ...
Page 88: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 88 of 110 ...
Page 89: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 89 of 110 ...
Page 90: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 90 of 110 ...
Page 91: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 91 of 110 ...
Page 92: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 92 of 110 ...
Page 93: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 93 of 110 ...
Page 95: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 95 of 110 ...
Page 96: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 96 of 110 5 4 PIPE SIZE DATA ...
Page 97: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 97 of 110 ...
Page 98: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 98 of 110 ...
Page 99: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 99 of 110 ...
Page 102: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 102 of 110 REGISTER NAME VARIABLE NUMBER FORMAT NOTE ...
Page 103: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 103 of 110 REGISTER NAME VARIABLE NUMBER FORMAT NOTE ...
Page 107: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 107 of 110 ...
Page 108: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 108 of 110 ...
Page 109: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 109 of 110 ...