HOW COULD THE CURRENT OUTPUT BE CALIBRATED?
Generally, the current outputs is factory calibrated with high precision and is it rarely
necessary to modify it. If during the simulation in M58 the device would measure a
value different from the foreseen one, please proceed as follows:
➢
Digit
MENU
/-
0
ENT
and the it will display
H a r d w a r e A d j u s t i n g
> _
Digit the code
4
2
1
3
0
6
8
and press
ENT
and the device will display:
A O H a r d w a r e A d j u s t i n g M 1
P r e s s E N T W h e n R e a d y
Press
ENT
again and the device will display:
A O C a l i b r a t i o n
[ M . 1
4 m A = = > 9 8 8 2 . 2
It is suggested to take note of the displayed value (+/-) before changing it. Press the
arrow keys and you will be able to change the value, it is also possible to digit it directly.
When changed, press
ENT
again and the device will display:
A O C a l i b r a t i o n
[ M . 1
2 0 m A = = > 5 0 1 1 2
It is suggested to take note of the displayed value (+/-) before changing it. Press the
arrow keys and you will be able to change the value, it is also possible to digit it directly.
When changed, press
ENT
again and the device will display:
A O H a r d w a r e A d j u s t i n g M . 1
P r e s s E N T W h e n R e a d y
The changed value has been saved into the EEPROM (an internal memory which could be
deleted only electrically and which keeps the data saved for almost 10 years).
Summary of Contents for TTFM100-F18-NG
Page 23: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 23 of 110 Or Or ...
Page 30: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 30 of 110 V mounting method N mounting method ...
Page 78: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 78 of 110 ...
Page 79: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 79 of 110 ...
Page 80: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 80 of 110 5 2 SOUND SPEEDS IN FLUIDS ...
Page 81: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 81 of 110 ...
Page 82: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 82 of 110 ...
Page 83: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 83 of 110 ...
Page 84: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 84 of 110 ...
Page 85: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 85 of 110 ...
Page 86: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 86 of 110 ...
Page 87: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 87 of 110 ...
Page 88: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 88 of 110 ...
Page 89: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 89 of 110 ...
Page 90: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 90 of 110 ...
Page 91: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 91 of 110 ...
Page 92: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 92 of 110 ...
Page 93: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 93 of 110 ...
Page 95: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 95 of 110 ...
Page 96: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 96 of 110 5 4 PIPE SIZE DATA ...
Page 97: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 97 of 110 ...
Page 98: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 98 of 110 ...
Page 99: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 99 of 110 ...
Page 102: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 102 of 110 REGISTER NAME VARIABLE NUMBER FORMAT NOTE ...
Page 103: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 103 of 110 REGISTER NAME VARIABLE NUMBER FORMAT NOTE ...
Page 107: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 107 of 110 ...
Page 108: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 108 of 110 ...
Page 109: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 109 of 110 ...