DIAGNOSTICS MENU ANALYSIS
➢
Digit
MENU
9
0
and the device will display
S t r e n t h + Q u a l i t y [ 9 0
U P : x x . x D N : x x . x Q = x x . x
This window displays the strength and the quality of the signal with numbers from 00.0 to
99.9.
00.0 means that there is no signal, 99.9 means that the signal is at its maximum value.
In normal working conditions the strength signal should be around 60.0.
During installation, please pay attention that the quality and strength have to be at their
max values.
➢
Digit
MENU
9
1
and the device will display
T O M / T O S * 1 0 0
[ 9 1
1 0 0 , 2 5 %
This window displays the relationship between the transit time and the measured time. In
normal working conditions, it should be 100 +/- 3%. Bigger differences indicate an error
in mounting or the setting of wrong parameters.
➢
Digit
MENU
9
2
and the device will display
F l u i d S o u n d V e l o c i t y
1 4 8 2 . 5 6 m / s
This window displays the speed of sound through the fluid. In normal working conditions,
the value should be similar to the one indicated in menu M21. Big differences indicate an
error during the sensors mounting or an error in M21.
➢
Digit
MENU
9
3
and the device will display
T o t a l T i m e , D e l t a T i m e
6 2 3 , 8 0 u S ,
2 4 2 , 1 2 n S
This window displays the total transit time and the UP-DOWN difference between transit
times. The values give information about the installation. In normal working conditions the
difference is less than 10%.
If the pipe diameter is small or the fluid velocity is very low, the difference might be a little
bit higher. If the difference between the flow rate and the speed is too high, the signal is
very bad
. The causes could be: pipe’s features, wrong mounting, wrong parameter
s
settings.
Summary of Contents for TTFM100-F18-NG
Page 23: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 23 of 110 Or Or ...
Page 30: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 30 of 110 V mounting method N mounting method ...
Page 78: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 78 of 110 ...
Page 79: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 79 of 110 ...
Page 80: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 80 of 110 5 2 SOUND SPEEDS IN FLUIDS ...
Page 81: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 81 of 110 ...
Page 82: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 82 of 110 ...
Page 83: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 83 of 110 ...
Page 84: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 84 of 110 ...
Page 85: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 85 of 110 ...
Page 86: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 86 of 110 ...
Page 87: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 87 of 110 ...
Page 88: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 88 of 110 ...
Page 89: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 89 of 110 ...
Page 90: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 90 of 110 ...
Page 91: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 91 of 110 ...
Page 92: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 92 of 110 ...
Page 93: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 93 of 110 ...
Page 95: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 95 of 110 ...
Page 96: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 96 of 110 5 4 PIPE SIZE DATA ...
Page 97: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 97 of 110 ...
Page 98: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 98 of 110 ...
Page 99: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 99 of 110 ...
Page 102: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 102 of 110 REGISTER NAME VARIABLE NUMBER FORMAT NOTE ...
Page 103: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 103 of 110 REGISTER NAME VARIABLE NUMBER FORMAT NOTE ...
Page 107: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 107 of 110 ...
Page 108: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 108 of 110 ...
Page 109: ...Mod BM43 COM Vers 1 0 16 11 2005 Page 109 of 110 ...